Diagnostic performance of a point of care high-sensitivity cardiac troponin I assay and single measurement evaluation to rule out and rule in acute coronary syndrome
Author:
De Iuliis Vincenzo1ORCID, Gabriele Anna Rita2, De Santis Francesca1, De Rugeriis Roberta1, Di Quinzio Luigi2, Aloisi Steeve2, Rosati Anna Carla3, Benvenuto Manuela3, Fabiani Donatello3, Chiatamone Ranieri Sofia1
Affiliation:
1. Department of Clinical Pathology , 92708 G. Mazzini Civil Hospital , Teramo , Italy 2. Emergency Department , 92708 G. Mazzini Civil Hospital , Teramo , Italy 3. Cardiology Division , 92708 G. Mazzini Civil Hospital , Teramo , Italy
Abstract
Abstract
Objectives
About 10 million individuals in USA presented annually in the emergency department (ED) with chest pain or with signs and symptoms of acute coronary syndrome (ACS). The advent of point of care (POC) devices, able to measure high sensitivity troponin, are a very interesting tool in the ED setting for its rapid turnaround time (<10 min).
Methods
The present study evaluates the diagnostic performance of the Atellica VTLi (Siemens) in real life setting using the clinical data derived from integrated diagnoses of emergency room staff and cardiologist and in comparison with standard laboratory hs-cTnT assay (Cobas 8000, Elecsys, Roche). 966 patients admitted to the emergency department of “G. Mazzini Hospital” in Teramo, Italy, from July 27, 2022, through June 09, 2023, were enrolled.
Results
The diagnostic performance of POC hs-cTnI was evaluated. An appropriate POC hs-cTnI threshold values <4 ng/L supplied a sensitivity and an NPV of 100 % (95 % CI: 99.5–100) in order to achieve rapid rule out for MI through a single measurement at patient presentation in the ED. Furthermore, a derivation POC hs-cTnI concentration >54 ng/L provided a specificity of 97.2 % (95 % CI: 95.9–98.1) and a PPV of 43.5 % (95 % CI: 40.3–46.7) for ruling in MI.
Conclusions
This platform showed comparable diagnostic performance for myocardial infarction to the central laboratory. Our data suggest the possible use of the Atellica VTLi hs-cTnI POC assay either in emergency department of urban medical centre, either in rural hospital for triage and patient management.
Publisher
Walter de Gruyter GmbH
Reference27 articles.
1. Go, AS, Mozaffarian, D, Roger, VL, Benjamin, EJ, Berry, JD, Blaha, MJ, et al.. Heart disease and stroke statistics-2014 update: a report from the American Heart Association. Circulation 2014;129:e28–92. https://doi.org/10.1161/01.cir.0000441139.02102.80. 2. Tomaszewski, CA, Nestler, D, Shah, KH, Sudhir, A, Brown, MD. Clinical policy: critical issues in the evaluation and management of emergency department patients with suspected non-ST-elevation acute coronary syndromes. Ann Emerg Med 2018;72:e65-106. https://doi.org/10.1016/j.annemergmed.2018.07.045. 3. Collet, JP, Thiele, H, Barbato, E, Barthélémy, O, Bauersachs, J, Bhatt, DL, et al.. 2020 ESC guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation. Eur Heart J 2021;42:1289–367. https://doi.org/10.1093/eurheartj/ehaa575. 4. Chew, DP, Lambrakis, K, Blyth, A, Seshadri, A, Edmonds, MJR, Briffa, T, et al.. A randomized trial of a 1-hour troponin T protocol in suspected acute coronary syndromes: the rapid assessment of possible acute coronary syndrome in the emergency department with high-sensitivity troponin T study (RAPID-TnT). Circulation 2019;140:1543–56. https://doi.org/10.1161/CIRCULATIONAHA.119.042891. 5. Westwood, ME, Armstrong, N, Worthy, G, Fayter, D, Ramaekers, BLT, Grimm, S, et al.. Optimizing the use of high-sensitivity troponin assays for the early rule-out of myocardial infarction in patients presenting with chest pain: a systematic review. Clin Chem 2021;67:237–44. https://doi.org/10.1093/clinchem/hvaa280.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|