Evaluation of serum electrolytes measurement through the 6-year trueness verification program in China
Author:
Yan Ying1, Pu Yungang1, Zeng Jie1, Zhang Tianjiao1, Zhou Weiyan1, Zhang Jiangtao1, Wang Jing1, Zhang Chao1, Chen Wenxiang1, Zhang Chuanbao1
Affiliation:
1. National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Beijing Hospital, National , Center of Gerontology, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences , Beijing , P.R. China
Abstract
Abstract
Objectives
The National Center for Clinical Laboratories (NCCL) in China initiated a serum electrolyte trueness verification (ETV) program in 2014 for measurement standardization.
Methods
Every year, two levels of fresh frozen commutable serum samples determined by inductively coupled plasma mass spectrometry (ICP-MS) reference methods were transported to participating clinical laboratories for the measurement of sodium, potassium, calcium and magnesium. Both samples were measured 15 times in 3 days, and the mean values and coefficient variations (CVs) were calculated from the results. The tolerance limits of trueness (bias), precision (CV) and accuracy (TE) based on the biological variation database were used as the evaluation criteria. The overall trend of the ETV program over 6 years was surveyed by calculating the pass rates of the participating laboratories. The mean bias, inter-laboratory CV, and TE of all laboratory results were analysed. Furthermore, homogeneous and heterogeneous systems were compared, and the bias and CV results of mainstream analysis systems were analysed.
Results
Pass rates of the three quality specifications increased, and the overall mean bias and inter-laboratory CVs decreased. The homogeneous system was superior to the heterogeneous system for calcium and magnesium measurements. For sodium, potassium, calcium and magnesium, the minimum bias corresponded to Hitachi, Siemens, Beckman AU and Roche, respectively. For inter-laboratory robust CVs, no obvious differences were observed between each peer group.
Conclusions
The commutable ETV materials assigned via reference methods can evaluate the accuracy and reproducibility of an individual laboratory and the calibration traceability and uniformity between laboratories for measurements.
Funder
National Key Technology R&D Program of China
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference30 articles.
1. Burtis, CA, Ashwood, ER, Bruns, DE, editors. Fundamentals of clinical chemistry, 6th ed. St. Louis, MO: Saunders; 2007:432 p. 2. Scott, MG, LeGrys, VA. Electrolytes and blood gases. In: Burtis, CA, Ashwood, ER, Bruns, DE, editors. Tietz fundamentals of clinical chemistry, 6th ed. Missouri: Saunders Press; 2008:431–5 pp. 3. Brady, H, Ryan, M, Horgan, J. Magnesium: the forgotten cation. Ir Med J 1987;80:250–3. 3312087. 4. Gowenlock, AH, McMurray, JR, McLauchlan, DH, editors. Varley’s practical clinical biochemistry, 6th ed. London: Heinemann Medical Books; 1988. 5. Belk, WP, Sunderman, FW. A survey of the accuracy of chemical analyses in clinical laboratories. Am J Clin Pathol 1947;17:853–61. https://doi.org/10.1093/ajcp/17.11.853.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|