Evaluation of four automated clinical analyzers for the determination of total 25(OH)D in comparison to a certified LC-MS/MS
Author:
Favresse Julien12, Fangazio Marco3, Cotton Frédéric3, Wolff Fleur3
Affiliation:
1. Department of Laboratory Medicine , Clinique St-Luc Bouge , Namur , Belgium 2. Department of Pharmacy , NAmur Research Institute for LIfe Sciences, University of Namur , Namur , Belgium 3. Department of Clinical Chemistry , LHUB-ULB, Université libre de Bruxelles , Brussels , Belgium
Abstract
Abstract
Objectives
The aim of this study was to compare the results of five methods for the determination of total 25(OH)D. For that purpose, two mass spectrometry and three immunoassay methods were used.
Methods
A total of 124 serum samples were analyzed on five different methods (i.e., a reference LC-MS/MS, Cascadion, Lumipulse, Roche Elecsys II and Roche Elecsys III). Analytical performance against LC-MS/MS was evaluated and compared to the Milan models 1 (analytical performance based on the clinical outcome using thresholds of 12, 20 and 30 ng/mL) and 2 (analytical performance based on biological variation). Additionally, imprecision studies and accuracy using NIST SRM972a samples were carried out.
Results
Compared to the reference LC-MS/MS method, the Lumipulse and the Roche Elecsys III assays reached the optimal criterion for bias, while the Cascadion met the desirable one. The Roche Elecsys II was not able to reach the minimal criteria. The proportion of correctly classified patients was higher using the Cascadion (95.2%) compared to the three immunoassays. In addition to its better precision, the Cascadion was not impacted by a high concentration of 3-epi-25(OH)D3 compared to the three immunoassays.
Conclusions
Compared to the LC-MS/MS reference method, the Cascadion presented the highest level of concordance at medical decision cut-offs for total 25(OH)D and reached the desirable specification for bias. Moreover, the presence of 3-epi-25(OH)D3 in enriched samples was only problematic in immunoassay methods, and especially considering Roche Elecsys methods. The release of performant fully automated mass spectrometry assays with high throughput might therefore facilitate the wide scale adoption of LC-MS/MS, even in non-specialized clinical laboratories.
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference25 articles.
1. Rifai, N, Horvath, A, Wittwer, C. Tietz textbook of clinical chemistry and molecular diagnostics, 6th ed. St. Louis, Missouri: Elsevier; 2018. 2. Bischoff-Ferrari, HA, Dawson-Hughes, B, Stocklin, E, Sidelnikov, E, Willett, WC, Edel, JO, et al.. Oral supplementation with 25(OH)D3 versus vitamin D3: effects on 25(OH)D levels, lower extremity function, blood pressure, and markers of innate immunity. J Bone Miner Res 2012;27:160–9. https://doi.org/10.1002/jbmr.551. 3. Dawson-Hughes, B, Mithal, A, Bonjour, JP, Boonen, S, Burckhardt, P, Fuleihan, GE, et al.. IOF position statement: vitamin D recommendations for older adults. Osteoporos Int 2010;21:1151–4. https://doi.org/10.1007/s00198-010-1285-3. 4. Holick, MF, Binkley, NC, Bischoff-Ferrari, HA, Gordon, CM, Hanley, DA, Heaney, RP, et al.. Evaluation, treatment, and prevention of vitamin D deficiency: an endocrine society clinical practice guideline. J Clin Endocrinol Metab 2011;96:1911–30. https://doi.org/10.1210/jc.2011-0385. 5. Pittas, AG, Sun, Q, Manson, JE, Dawson-Hughes, B, Hu, FB. Plasma 25-hydroxyvitamin D concentration and risk of incident type 2 diabetes in women. Diabetes Care 2010;33:2021–3. https://doi.org/10.2337/dc10-0790.