Evaluation of four automated clinical analyzers for the determination of total 25(OH)D in comparison to a certified LC-MS/MS

Author:

Favresse Julien12,Fangazio Marco3,Cotton Frédéric3,Wolff Fleur3

Affiliation:

1. Department of Laboratory Medicine , Clinique St-Luc Bouge , Namur , Belgium

2. Department of Pharmacy , NAmur Research Institute for LIfe Sciences, University of Namur , Namur , Belgium

3. Department of Clinical Chemistry , LHUB-ULB, Université libre de Bruxelles , Brussels , Belgium

Abstract

Abstract Objectives The aim of this study was to compare the results of five methods for the determination of total 25(OH)D. For that purpose, two mass spectrometry and three immunoassay methods were used. Methods A total of 124 serum samples were analyzed on five different methods (i.e., a reference LC-MS/MS, Cascadion, Lumipulse, Roche Elecsys II and Roche Elecsys III). Analytical performance against LC-MS/MS was evaluated and compared to the Milan models 1 (analytical performance based on the clinical outcome using thresholds of 12, 20 and 30 ng/mL) and 2 (analytical performance based on biological variation). Additionally, imprecision studies and accuracy using NIST SRM972a samples were carried out. Results Compared to the reference LC-MS/MS method, the Lumipulse and the Roche Elecsys III assays reached the optimal criterion for bias, while the Cascadion met the desirable one. The Roche Elecsys II was not able to reach the minimal criteria. The proportion of correctly classified patients was higher using the Cascadion (95.2%) compared to the three immunoassays. In addition to its better precision, the Cascadion was not impacted by a high concentration of 3-epi-25(OH)D3 compared to the three immunoassays. Conclusions Compared to the LC-MS/MS reference method, the Cascadion presented the highest level of concordance at medical decision cut-offs for total 25(OH)D and reached the desirable specification for bias. Moreover, the presence of 3-epi-25(OH)D3 in enriched samples was only problematic in immunoassay methods, and especially considering Roche Elecsys methods. The release of performant fully automated mass spectrometry assays with high throughput might therefore facilitate the wide scale adoption of LC-MS/MS, even in non-specialized clinical laboratories.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3