Innovations in HbA1c analysis: finding the balance between speed and accuracy. An investigation of a potential new Secondary Reference Measurement Procedure for the IFCC

Author:

Lenters-Westra Erna12,English Emma3

Affiliation:

1. Department of Clinical Chemistry , Isala , Zwolle , The Netherlands

2. European Reference Laboratory for Glycohemoglobin , Zwolle , The Netherlands

3. Institute of Continuing Education , University of Cambridge , Cambridge , Cambridgeshire , UK

Abstract

Abstract Objectives The escalating prevalence of diabetes worldwide has resulted in a dramatic increase in the number of people who need testing, which in turn necessitates faster HbA1c measurement. The Tosoh GR01 addresses the need for fast turnaround times of whilst offering pragmatic steps to maintain result accuracy in a single instrument by offering two distinct operating modes: Short Mode (SM) and Long Mode (LM). The aim of this study was to evaluate all relevant aspects of the performance of the Tosoh GR01 with a view to accepting the instrument as a future Secondary Reference Measurement Procedure (SRMP) for the IFCC. Methods Certified Clinical & Laboratory Standards Institute (CLSI) Evaluation Protocols (EP) were used to evaluate precision (EP-5), accuracy (EP-9), linearity (EP-6), carry-over (EP-10) and the effect of hemoglobin variants and other potential interferences. Results Both modes demonstrated CVs <0.6 % in SI units and <0.4 % in NGSP units at 46 mmol/mol (6.4 %) and 75 mmol/mol (9.0 %) and passed both National Glycohemoglobin Standardization Program (NGSP) and International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) certification procedures when compared with 7 IFCC Certified Secondary Reference Measurement Procedures (SRMP). Sigma for both modes was >6 when using the results of EP-5 and EP-9 at an HbA1c concentration of 50 mmol/mol (6.7 %). Neither mode showed any interference with common Hb-variants except for HbAE when HbA1c was >65 mmol/mol. In the SM HbAS, HbAD and HbAC were recognized but no result was reported. Conclusions There is a good balance between speed and accuracy for determining HbA1c with the Tosoh GR01 in both analytical modes and the device is suitable for use as an IFCC SRMP.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3