Lot variation and inter-device differences contribute to poor analytical performance of the DCA Vantage™ HbA1c POCT instrument in a true clinical setting

Author:

Abildgaard Anders12ORCID,Knudsen Cindy Søndersø3,Bjerg Lise Nørkjær4,Lund Sten2,Støy Julie2

Affiliation:

1. Department of Clinical Biochemistry , Regional Hospital Horsens , Horsens , Denmark

2. Steno Diabetes Center Aarhus , Aarhus University Hospital , Aarhus N , Denmark

3. Department of Clinical Biochemistry , Aarhus University Hospital , Aarhus N , Denmark

4. Department of Clinical Biochemistry , Regional Hospital Central Jutland , Viborg , Denmark

Abstract

Abstract Objectives The glycated haemoglobin fraction A1c (HbA1c) is widely used in the management of diabetes mellitus, and the Siemens DCA Vantage™ point-of-care testing (POCT) instrument offers rapid HbA1c results even far from a clinical laboratory. However, the analytical performance has been questioned, and not much is known about effects of changing reagent lot, instrument and operator. We therefore compared the analytical performance of the DCA Vantage™ with established routine methods (Tosoh G8/G11 ion exchange HPLC) in a true clinical setting at two Danish hospitals. Methods We extracted all routine clinical HbA1c results incidentally drawn from the same patient within 48 h (n=960 pairs) and evaluated the effect of reagent lot, operator and instrument. We also performed a prospective method comparison in our diabetes out-patient clinic (n=97). Results The critical difference (CD) between two POCT results varied between 5.14 and 6.61 mmol/mol (0.47–0.55%), and the analytical imprecision of the DCA Vantage™ (CVA) was >3%. Significant effect of reagent lot and inter-instrument differences were found, whereas no effect of operator was seen. Conclusions The DCA Vantage™ HbA1c analysis does not fulfil the prevailing analytical performance specifications, but rigorous validation of new reagent lots and continuous recalibration of instruments may potentially improve the precision substantially. Our findings, therefore, clearly emphasise the necessity of a close collaboration between clinicians and laboratory professionals in the POCT field. Finally, POCT HbA1c results should always be interpreted together with other measures of glycaemic control to avoid inappropriate change of patient treatments due to measurement uncertainty.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3