Calibration of insulin pumps based on discrete doses at given cycle times

Author:

Bissig Hugo1,Büker Oliver2ORCID,Stolt Krister2,Batista Elsa3,Afonso Joana4,Zagnoni Michele5,Vroman Rozan5,Kjeldsen Henrik6,Niemann Anders6,Schroeter Joerg7

Affiliation:

1. Physics Department , Federal Institute of Metrology METAS , Bern-Wabern , Switzerland

2. RISE Research Institutes of Sweden , Borås , Sweden

3. Metrology Department , IPQ , Caparica , Portugal

4. UNIDEMI, NOVA School of Science and Technology , Caparica , Portugal

5. Department of Electrical and Electronic Engineering , University of Strathclyde (UoS) , Glasgow , UK

6. Danish Technological Institute (DTI) , Aarhus C , Denmark

7. Medical Sensors and Devices Lab , Technische Hochschule Lübeck (THL) , Lübeck , Germany

Abstract

Abstract One application in the medical treatment at very small flow rates is the usage of an Insulin pump that delivers doses of insulin at constant cycle times for a specific basal rate as quasi-continuous insulin delivery, which is an important cornerstone in diabetes management. The calibration of these basal rates are performed by either gravimetric or optical methods, which have been developed within the European Metrology Program for Innovation and Research (EMPIR) Joint Research Project (JRP) 18HLT08 Metrology for drug delivery II (MeDDII). These measurement techniques are described in this paper, and an improved approach of the analytical procedure given in the standard IEC 60601-2-24:2012 for determining the discrete doses and the corresponding basal rates is discussed in detail. These improvements allow detailed follow up of dose cycle time and delivered doses as a function of time to identify some artefacts of the measurement method or malfunctioning of the insulin pump. Moreover, the calibration results of different basal rates and bolus deliveries for the gravimetric and the optical methods are also presented. Some analysis issues that should be addressed to prevent misinterpreting of the calibration results are discussed. One of the main issues is the average over a period of time which is an integer multiple of the cycle time to determine the basal rate with the analytical methods described in this paper.

Funder

European Metrology Programme for Innovation and Research

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Evidence and Diabetes Technology: A Catch-22?;Journal of Diabetes Science and Technology;2023-04-27

2. Medical flow and dosing measurement metrology in drug delivery;Biomedical Engineering / Biomedizinische Technik;2022-12-26

3. Metrology in health: challenges and solutions in infusion therapy and diagnostics;Biomedical Engineering / Biomedizinische Technik;2022-11-10

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3