Affiliation:
1. Section of Clinical Biochemistry , 19051 University of Verona , Verona , Italy
Abstract
Abstract
Objectives
The preanalytical phase in clinical laboratory diagnostics is currently receiving more and more attention. This term describes one part of actions and aspects of the “brain-to-brain cycle” of the medical laboratory diagnostic procedure that take place before the analytical phase. However, the preanalytical activities, the handling of unsuitable samples and the reporting procedures are neither fully standardized nor harmonized worldwide. The influence of the properties of the blood collection needle must be acknowledged. In this work, we focused on the investigation of the internal structure and size of standardized 21G blood collection needles.
Methods
All parameters were measured with a scanning electron microscope using a Jeol model JSM-6000PLUS. Our.
Results
The obtained data shows that the internal surfaces of the needles vary greatly from manufacturer to manufacturer (by around 35 %), and this may play an important role in influencing blood flow and even the risk of blood cell injury (especially hemolysis) during blood drawing.
Conclusions
The differential actual needle diameters can vary greatly between needle manufactures and this variety may have a significant impact on laboratory values and may also lead to specimen rejection.
Reference15 articles.
1. Lima-Oliveira, G, Lippi, G, Salvagno, GL, Picheth, G, Guidi, GC. Laboratory diagnostics and quality of blood collection. J Med Biochem 2015;34:288. https://doi.org/10.2478/jomb-2014-0043.
2. International Organization for Standardization. ISO 15189. Medical laboratories particular requirements for quality and competence; 2007.
3. Guder, WG. History of the preanalytical phase: a personal view. Biochem Med 2014;24:25–30. https://doi.org/10.11613/bm.2014.005.
4. Simundic, A, Lippi, G. Preanalytical phase – a continuous challenge for laboratory professionals. Biochem Med 2012;22:145–9. https://doi.org/10.11613/bm.2012.017.
5. Clinical Laboratory Standards Institute. CLSI GP41. Collection of Diagnostic Venous Blood Specimens; 2017.