Rayana Mohammed C. Ben,Burnett Robert W.,Covington Arthur K.,D'Orazio Paul,Fogh-Andersen Niels,Jacobs Ellis,Kataky Ritu,Külpmann Wolf R.,Kuwa Katsuhiko,Larsson Lasse,Lewenstam Andrzej,Maas Anton H.J.,Mager Gerhard,Naskalski Jerzy W.,Okorodudu Anthony O.,Ritter Christoph,St John Andrew
Abstract
AbstractThe proposed recommendation for measuring and reporting chloride in undiluted plasma or blood by ion-selective electrodes (ISEs) will provide results that are identical to chloride concentrations measured by coulometry for standardized normal plasma or blood samples. It is applicable to all current ISEs dedicated to chloride measurement in undiluted samples that meet the requirements. However, in samples with reduced water concentration, results by coulometry are lower than by ion-selective electrode due to volume displacement. The quantity measured by this standardized ISE procedure is called the ionized chloride concentration. It may be clinically more relevant than the chloride concentration as determined by coulometry, photometry or by ISE after dilution of the sample.
Subject
Biochemistry, medical,Clinical Biochemistry,General Medicine
Reference8 articles.
1. Donnan exclusion failure in low anion site density membranes containing valinomycin;Buck;J Electroanal Chem,1987
2. de Comparison of the precision of seven analytical methods for the concentration in human serum and urine;Jong;Clin Chim Acta,1987
3. The Donnan exclusion failure of neutral ionophore - based ion - selective electrodes studied by optical second - harmonic generation;Yajima;Anal Chem,1997
4. pH determination of body fluids with a micro glass electrode and a saturated KCl bridge in the cell;Maas;Clin Chim Acta,1970
5. Threshold ionic site concentrations required for Nernstian potentiometric responses of neutral ionophore - incorporated ion - selective liquid membrane;Tohda;Anal Sci,2001