Guidelines for sampling, measuring and reporting ionized magnesium in undiluted serum, plasma or blood: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC): IFCC Scientific Division, Committee on Point of Care Testing

Author:

Ben Rayana Mohammed C.,Burnett Robert W.,Covington Arthur K.,D'Orazio Paul,Fogh-Andersen Niels,Jacobs Ellis,Külpmann Wolf R.,Kuwa Katsuhiko,Larsson Lasse,Lewenstam Andrzej,Maas Anton H. J.,Mager Gerhard,Naskalski Jerzy H. J.,Okorodudu Anthony O.,Ritter Christoph,St John Andrew

Abstract

AbstractAll analyzers with ion-selective electrodes for ionized magnesium (iMg) should yield comparable and unbiased results. The prerequisite to achieve this goal is to reach consensus on sampling, measurement and reporting. The recommended guidelines for sampling, measurement and reporting iMg in plasma (“plasma” refers to circulating plasma and the forms in which it is sampled: the plasma phase of anticoagulated whole blood, plasma separated from blood cells, or serum) or blood, referring to the substance concentration of iMg in the calibrants, will provide results for iMg that are approximately 3% greater than its true concentration, and 4% less than its true molality. Binding of magnesium to proteins and ligands in plasma and blood is pH-dependent. Therefore, pH should be simultaneously measured to allow adjustment of iMg concentration to pH7.4. The substance concentration of iMg may be physiologically and consequently clinically more relevant than the substance concentration of total magnesium.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

Reference34 articles.

1. editors Unique magnesium - sensitive ion selective electrodes;Altura;Scand Clin Lab Invest,1994

2. Potentiometric response of magnesium - selective membrane electrode in the presence of nonionic surfactants;Malinowska;Anal Chim Acta,1999

3. Application of Nernst - Planck and Poisson equations for interpretation of liquid - junction and membrane potentials in real - time and space domains;Sokalski;Electrochem Commun,2001

4. A method for the measurement of ionized magnesium;Filos;Lab Med,1995

5. Measurement of ionized magnesium with neutral carrier based ISEs Progress and results with the AVL magnesium analyser;Marsoner;Scand Clin Lab Invest,1994

Cited by 11 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献