Author:
Batista Elsa,Almeida Nelson,Furtado Andreia,Filipe Eduarda,Sousa Luis,Martins Rui,Lucas Peter,Petter Harm Tido,Snijder Roland,Timmerman Annemoon
Abstract
AbstractFor critical drug delivery, it is important to have a constant and well-known infusion rate delivered by the complete infusion set-up (pump, tubing, and accessories). Therefore, various drug delivery devices and accessories were tested in this article in terms of their infusion accuracy, start-up delay, response time, and dependency on the viscosity. These measurements were performed as part of the European funded research project MeDD. The obtained results show that the infusion accuracy of the devices is flow rate and accessory depended, especially for low flow rates. Viscosity does not have a significant impact on the flow rate accuracy.
Reference30 articles.
1. considerations and initial validation of a liquid microflow calibration setup using parallel operated syringe pumps Publishing;Melvad;Measurement Science,2010
2. Technical data infusomat volumetric pump and perfusor space available from http www space bbraun com documents Space System Technical Data l pdf
3. Impact of syringe size on the performance of infusion pumps at low flow rates;Schmidt;Crit Care Med,2010
4. IEC ISO Evaluation of measurement data Supplement to the guide to the expression of uncertainty in measurement propagation of distributions using a Monte Carlo method Joint committee for guides in metrology;BIPM,2008
5. Primary standards for measuring flow rates from min to min gravimetric principle;Bissig;Biomed Eng Biomed Tech