The Intra-Cochlear Impedance-Matrix (IIM) test for the Nucleus® cochlear implant

Author:

Hey MatthiasORCID,Böhnke Britta,Dillier Norbert,Hoppe Ulrich,Eskilsson Gunnar,Löwgren Karolina,Cullington Helen,Mauch Herbert,Müller-Deile Joachim

Abstract

AbstractTo describe the principles and operation of a new telemetry-based function test for the NucleusThe IIM test measures bipolar impedances between all electrode pairs and employs a normalization procedure based on common ground impedances in order to identify abnormal current paths among electrodes. Six European clinics collected IIM data from a total of 192 devices.Reproducibility was high between initial and repeat measurements. The normative analysis demonstrated narrow ranges among devices after normalization of impedance data. The IIM is able to identify abnormal current paths that are not evident from standard impedance telemetry and may otherwise only be found utilising average electrode voltage measurements (AEV).The IIM test was found to be straightforward to perform clinically and demonstrated reproducible data with narrow ranges in normally-functioning devices. Because this test uses a very low stimulation level the IIM test is well suited for children or multiply handicapped CI users who cannot reliably report on their auditory percepts. The new algorithms show potential to improve implant integrity testing capabilities if implemented in future clinical software.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering

Reference86 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3