Routine coagulation testing in Vacutainer® Citrate Plus tubes filled at minimum or optimal volume

Author:

Balboni Fiamma1,Barbui Stefania2,Gallo Marco3,Berardi Margherita1,Vezzosi Marco1,Lippi Giuseppe4

Affiliation:

1. Service of Laboratory Medicine, Istituto Fiorentino Cura e Assistenza (IFCA) , Florence , Italy

2. Pre-hospital Admission Service, Istituto Fiorentino Cura e Assistenza (IFCA) , Florence , Italy

3. Hemodialysis Unit, Istituto Fiorentino Cura e Assistenza (IFCA) , Florence , Italy

4. Section of Clinical Biochemistry , University Hospital of Verona , Piazzale LA Scuro , 37134 Verona , Italy

Abstract

Abstract Background Filling of citrate tubes with appropriate amount of blood is essential for obtaining reliable results of coagulation testing. This study aimed to verify whether results of routine coagulation tests are comparable when the new Becton Dickinson Vacutainer® Citrate Plus tubes are filled at minimum or optimal volume. Methods The study population consisted of 133 patients (40 on oral anticoagulant therapy), who had blood collected for routine coagulation testing. Two sequential Vacutainer® Citrate Plus tubes of the same type and lot were drawn. The first tube was collected after a butterfly needle was inserted into the vein, so that the air in the tubing was aspirated into the tube before blood (minimum fill volume), whilst the second was drawn at optimal fill volume. Experiments were repeated using 2.7-mL (n = 86) and 1.8-mL (n = 47) tubes. Results Prothrombin time (PT) and fibrinogen values were slightly but significantly decreased in tubes with minimum than in those with optimal fill volume. The activated partial thromboplastin time (APTT) was slightly prolonged in tubes with minimum than in those with optimal fill volume, but the difference was not statistically significant. An identical trend was noted in separate analyses for the 2.7-mL and 1.8-mL tubes. Spearman’s correlations between the two fill volumes were always >0.94 and bias was always within the quality specifications. Conclusions Blood drawing into Vacutainer® Citrate Plus tubes at minimum fill volume does not clinically bias routine coagulation testing.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,Public Health, Environmental and Occupational Health,Health Policy,Medicine (miscellaneous)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3