Author:
Sesselmann Stefan,Hong Yutong,Schlemmer Frank,Wiendieck Kurt,Söder Stephan,Hussnaetter Isabell,Müller Lutz Arne,Forst Raimund,Wierer Thomas
Abstract
AbstractIn a previous radiostereometric analysis (RSA) of the Lubinus SP II (Link, Hamburg, Germany), which is one of the most often used cemented hip stems worldwide, our research group detected a very small but statistically significant distal migration of -0.03±0.17 mm 2 years after surgery compared to the postoperative radiograph. Maximum subsidence occurred between 6 and 12 months. The implant appeared to have stabilized after 2 years. The mean value of maximum total point motion (MTPM) was 0.99±0.69 mm, which was detected 2 years after surgery. The purpose of this study was to analyze the migration pattern and to verify the predictive value of short-term RSA of the Lubinus SP II stem after 10 years. After a follow-up of 5 and 10 years, 38 and 27 out of 100 patients remained available for further assessment, respectively. No statistically significant implant translation or rotation was found along or about the axes of the global coordinate system 5 and 10 years after surgery with respect to the postoperative radiograph. Furthermore, the MTPM was stable in both follow-up periods. The results suggest that the Lubinus SP II hip stem is still stable 10 years after surgery, supporting that determining prognosis by short-term RSA follow-up of 2 years could be an appropriate tool for appraisal of implant behavior 10 years after surgery.
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献