Affiliation:
1. Institute of Physics, College of General Education, Osaka University
2. Japan Electron Optics Laboratory
Abstract
A new type of mass spectrograph with large mass dispersion has been constructed. The distinguishing feature of this apparatus is that the mass dispersing action of a r-l magnetic field is utilized. The instrument has a r-1 magnetic field of 198.1° sector type (22 cm mean radius) as the dispersing field, a toroidal electric field of 118.7° sector type (30 cm mean radius) and a 30° uniform magnetic field (120 cm mean radius) as the focusing fields. The principle, the design and important parts of the apparatus are described. The dispersion on the photographic plate was estimated to be 14 cm for 1% mass difference and the maximum resolving power of about 500,000 was obtained.
Subject
Physical and Theoretical Chemistry,General Physics and Astronomy,Mathematical Physics
Cited by
21 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Prismatic mass analyzer with the conical achromatic prism and transaxial lenses;International Journal of Mass Spectrometry;2019-10
2. Hisashi Matsuda;The Encyclopedia of Mass Spectrometry;2015
3. Hisashi Matsuda;Journal of the Mass Spectrometry Society of Japan;2012
4. Mass Spectrometer Joint Development by Dr. Hisashi Matsuda ^|^amp; JEOL Ltd.;Journal of the Mass Spectrometry Society of Japan;2012
5. Struktur und Eigenschaften der Materie;Praktische Physik;1996