Abstract Objectives The results of external quality assurance schemes are evaluated against specifications generally based on biological variation (BV) data. This study was carried out to determine whether new BV values affected the level of compliance to specifications. Our secondary objective was to identify the conditions that would be compromised as a result of poor analytical performance in disease associated markers. Methods This study was based on the results of the SEQCML External Quality Assurance scheme for the 2015–2022 period. Deviation of the individual result from the target value was estimated. Additionally, we calculated the percentage of results that met the pre-established specification. Results In 97 of the 133 analytes, the level of compliance was maintained in 80–90 % of the results obtained in the two study periods. In 23 analytes, the level of compliance ranged from 51 to 79 % in the two study periods. In ALT, AST and sodium, the level of compliance was ≤50 % of the results obtained in the first study period, with sodium being the only analyte that maintained this poor level of compliance in the second study period. Conclusions The level of compliance to specifications remained independent from the specification used (SEQCML or EFLM) for the majority of the analytes. The results for sodium ion were below the target value, which may lead to misdiagnosis of hyponatremia. Non-compensated alkaline picrate methods overestimate creatinine, which may produce false information suggestive of kidney failure.

Publisher

Walter de Gruyter GmbH

Subject

Medical Laboratory Technology,Education,Medicine (miscellaneous)

Reference26 articles.

1. International Standardization Organization. Proficiency testing by interlaboratory comparisons – part 1: development and operation of proficiency testing schemes. Geneve: ISO/IEC Guide: 43; 1997.

2. IVDR 2017/746. Regulation (EU) 2017/746 of the European parliament and of the council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU; 2017.

3. Harel, D, Schumann, G, Libeer, JC. Guidelines for the requirements for the competence of EQAP organizers in medical laboratories. IFCC/EMD/C-AQ. IFCC; 2002. Available from: http://www.IFCC.org.

4. Plebani, M. Internal quality control and external quality assurance: a great pass opens the way to a brilliant future. Adv Lab Med 2022;3:215–17. https://doi.org/10.1515/almed-2022-0075.

5. Asociación española de certificación y normalización. Requisitos particulares para la calidad y la competencia. UNE-EN-ISO 15189:2007. Madrid: AENOR; 2013.