Evaluation of DiaSorin Liaison® calprotectin fecal assay adapted for pleural effusion

Author:

de Paz Poves Cristina1ORCID,Barneo-Caragol Clara1,Cillero Sánchez Ana Isabel2,Jimenez Mendiguchia Lucía3,Quirós Caso Covadonga1,Moreno Rodríguez María3,López González Francisco J.4,Prieto García Mᵃ Belén1ORCID

Affiliation:

1. Laboratory of Medicine, Department of Clinical Biochemistry , Hospital Universitario Central de Asturias , Oviedo , Asturias , Spain

2. Department of Clinical Biochemistry , Hospital Álvarez Buylla , Mieres , Asturias , Spain

3. Department of Clinical Biochemistry , Hospital Universitario San Agustín , Avilés , Asturias , Spain

4. Department of Pneumology , Hospital Universitario Central de Asturias , Oviedo , Asturias , Spain

Abstract

Abstract Objectives Calprotectin (CP) is a calcium and zinc binding protein that is widely measured on faecal samples but its determination in other biological fluids might be of interest. The aim of this work was to validate the measurement of CP in pleural fluid by chemiluminescence. Methods LIAISON®XL, a fully automated chemiluminescence analyzer, was used for CP quantification on pleural fluid. A validation protocol was designed using both quality control materials provided by the manufacturer and pools of pleural fluid samples. Stability, imprecision, bias, linearity, detection capability and carry over effect were evaluated. Results CP was stable on pleural fluid at least one week, under refrigerated conditions, and four weeks at −80 °C. The observed intra- and inter-day imprecision was between 2.2 and 6.49 %, with a negative bias under 5.51 %. The linearity of the method was verified up to 2,000 ng/mL. The LoQ for the assay was 48.52 ng/mL. A statistically significant carry-over effect was observed after measuring CP concentrations above the upper limit of linearity, but given the observed magnitude, a clinically relevant impact should not be expected. Conclusions DiaSorin Liaison® calprotectin assay allows reliable measurement of CP in pleural fluid.

Funder

Fundación José Luis Castaño

Publisher

Walter de Gruyter GmbH

Subject

Medical Laboratory Technology,Education,Medicine (miscellaneous)