Abstract
AbstractIn this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM). As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.
Reference13 articles.
1. Experimental design techniques for optimising measurement chain calibration;Betta;Measurement,2001
2. An improved intelligent calibration method for vortex flowmeter of the;Yi;Proc American Control Conference,2007
3. Optimum choice of measurement points for sensor calibration;Betta;Measurement,1996
4. Application of sensitivity analysis to the correction of static characteristics of a phase angle modulator Metrol;Sroka;Meas Syst,2011
5. Statistical measurement calibration based on state estimator results of the IEEE Power Engineering Society Transmission and Distribution Conference New Orleans;Moreno;Proc USA,1999
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献