The local technical validation of new plasma tube with a mechanical separator

Author:

Ucar Kamil Taha1ORCID,Aksoy Neval1,Erhan Belgin2,Bercik Inal Berrin1ORCID

Affiliation:

1. Department of Medical Biochemistry , Istanbul Gaziosmanpasa Taksim Education and Research Hospital , Istanbul , Turkey

2. Department of Physical Medicine and Rehabilitation , Istanbul Gaziosmanpasa Taksim Education and Research Hospital , Istanbul , Turkey

Abstract

AbstractBackgroundUnsuitable samples are common problem for laboratories. The blood collection tubes need to be validated or verified prior to their being used in the routine laboratory for reducing this situation.ObjectiveWe aim to compare the technical qualifications of routinely used BD Vacutainer®Serum Separator Tubes™ II Advance Plus with BD Vacutainer®Barricor™ LH Plasma Tubes for local technical validation.Materials and methodsApparently healthy 150 voluntary subjects were enrolled in the study. Samples were collected in two separated tubes by a single phlebotomist. Twelve quality indicators were used to compare these two different types of tubes for local technical validation. Differences (%) between them were calculated with the formula proposed by EFLM. In case of any difference of less than 1% for each indicator, the evaluated tube was considered as non-inferior.ResultsIndicators, such as tubes with physical defects, that fail to create vacuum, not properly fitting into the blood collection device, under filling (10%), cracked tubes, tubes exterior surface contaminated with blood, hemolysed specimens, including fibrin strand/mass in the sample, red blood cell adhesion, poor/incomplete barrier formation were found non-inferior in Barricor™ tubes. White particulate matter (WPM) was observed in 24.6% of Barricor™. Therefore, the last indicator (tubes including gel/foreign material/WPM in sample after centrifugation) was found inferior for Barricor™.ConclusionTechnical local validation studies should be encouraged in terms of quality management. It was thought that WPM would not cause any interference in a properly filled tube. In addition to, Barricor™ was also found to be technically acceptable when evaluated through using all other indicators.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,Molecular Biology,Biochemistry

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3