Abstract
Abstract
While the resolving power of drift tube ion mobility spectrometers has been studied and modelled in detail over the past decades, no comparable model exists for the signal-to-noise-ratio. In this work, we develop an analytical model for the signal-to-noise-ratio of a drift tube ion mobility spectrometer based on the same experimental parameters used for modelling the resolving power. The resulting holistic model agrees well with experimental results and allows simultaneously optimizing both resolving power and signal-to-noise-ratio. Especially, it reveals several unexpected relationships between experimental parameters. First, even though reduced initial ion packet widths result in fewer injected ions and reduced amplifier widths result in more noise, the resulting shift of the optimum operating point when reducing both simultaneously leads to a constant signal-to-noise-ratio. Second, there is no dependence of the signal-to-noise-ratio at the optimum operating point on the drift length, as again the resulting shift of the optimum operating point causes all effects to compensate each other.
Subject
Electrical and Electronic Engineering,Instrumentation
Reference34 articles.
1. G. A. Eiceman, Z. Karpas, H. H. Hill. Ion Mobility Spectrometry. CRC Press, 2013. ISBN: 1439859973.
2. A. E. Eiceman. Toward the chemical agent monitor: technologies and developments in England and the United States from 1965 to 1982. Int. J. Ion Mobil. Spec., 2020. DOI: 10.1007/s12127-019-00256-w.
3. H. Borsdorf, G. A. Eiceman. Ion mobility spectrometry: principles and applications. Applied Spectroscopy Reviews, 41(4):323–375, 2006. DOI: 10.1080/05704920600663469.
4. A. T. Kirk. Driftzeit-Ionenmobilitätsspektrometer mit hoher analytischer Leistungsfähigkeit - Modellierung, Entwurf und Umsetzung. Shaker, 2020. ISBN: 9783844074420.
5. A. T. Kirk, S. Zimmermann. Bradbury-Nielsen vs. Field switching shutters for high resolution drift tube ion mobility spectrometers. Int. J. Ion Mobil. Spec., 17(3-4):131–137, 2014. DOI: 10.1007/s12127-014-0153-9.
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献