Effects of Conventional Mechanical Ventilation Performed by Two Neonatal Ventilators on the Lung Functions of Rabbits with Meconium-Induced Acute Lung Injury

Author:

Mokra D1,Mikolka P1,Kosutova P1,Kolomaznik M1,Jurcek M2,Istona P2,Matasova K3,Zibolen M3,Calkovska A1

Affiliation:

1. Comenius University, Jessenius Faculty of Medicine in Martin, Biomedical Center and Department of Physiology, Martin, Slovakia

2. Chirana Stara Tura a.s., Slovakia

3. Comenius University, Jessenius Faculty of Medicine and University Hospital, Neonatology Clinic, Martin, Slovakia

Abstract

Abstract Severe meconium aspiration syndrome (MAS) in the neonates often requires a ventilatory support. As a method of choice, a conventional mechanical ventilation with small tidal volumes (VT<6 ml/kg) and appropriate ventilatory pressures is used. The purpose of this study was to assess the short-term effects of the small-volume CMV performed by two neonatal ventilators: Aura V (Chirana Stara Tura a.s., Slovakia) and SLE5000 (SLE Ltd., UK) on the lung functions of rabbits with experimentally-induced MAS and to estimate whether the newly developed neonatal version of the ventilator Aura V is suitable for ventilation of the animals with MAS. In the young rabbits, a model of MAS was induced by an intratracheal instillation of a suspension of neonatal meconium (4 ml/kg, 25 mg/ml). After creating the model of MAS, the animals were ventilated with small-volume CMV (frequency 50/min, VT <6 ml/kg, inspiration time 50 %, fraction of inspired oxygen 1.0, positive end-expiratory pressure 0.5 kPa, mean airway pressure 1.1 kPa) performed by ventilator Aura V (Aura group, n=7) or ventilator SLE5000 (SLE group, n=7) for additional 4 hours. One group of animals served as healthy non-ventilated controls (n=6). Blood gases, oxygenation indexes, ventilatory pressures, lung compliance, oxygen saturation and total and differential white blood cell (WBC) count were regularly determined. After euthanizing the animals, a left lung was saline-lavaged and total and differential counts of cells in the bronchoalveolar lavage (BAL) fluid were determined. A right lung was used for estimation of lung edema formation (expressed as a wet/dry weight ratio) and for analysis of concentrations of pro-inflammatory cytokines (IL-1β, IL-8, TNF). The cytokines were measured also in the blood plasma taken at the end of experiment. Meconium instillation seriously worsened the gas exchange and induced inflammation and lung edema formation. In the Aura group, slightly lower concentrations of cytokines were found and better gas exchange early after creating the MAS model was observed. However, there were no significant differences in the respiratory parameters between the ventilated groups at the end of experiment (P>0.05). Concluding, the newly developed neonatal version of the ventilator Aura V was found to be fully comparable to widely used neonatal ventilator SLE5000. Results provided by Aura V in CMV ventilation of rabbits with meconium-induced acute lung injury suggest its great potential also for future clinical use, i.e. for ventilation of the neonates with MAS.

Publisher

Walter de Gruyter GmbH

Subject

General Biochemistry, Genetics and Molecular Biology,General Nursing

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3