Affiliation:
1. Oral Technology , University of Bonn , Bonn , Germany
2. Department of Prosthetic Dentistry, Preclinical Education and Materials Science , Dental School, University of Bonn , Bonn , Germany
Abstract
Abstract
The aim of this study was to investigate experimentally the fatigue and stability of three bar materials with distal extension at the molar region and to numerically analyse the biomechanical properties of the bar materials connected to overdentures in a patient individual model. A milled bar was designed for the mandible on four implants in the canine and second premolar region. Three bar materials were investigated: titanium (Ti), cobalt chromium (CoCr), and polyetherketonketon (PEKK). Firstly, static and fatigue tests were performed based on EN ISO 14801 in a commercial permanent loading set-up. Unilateral axial force was applied on the distal extension of the bars. Secondly, numerical models were created. Different bar materials and loading scenarios were analysed. The static fracture limit of the three materials was 1,750 N, 780 N, 310 N for Ti, CoCr, and PEKK, respectively. The Wöhler curves showed comparable fatigue limits of 200 N, 160 N, and 150 N for titanium, CoCr, and PEKK, respectively. The stress at the distal extension was 2,600 MPa (Ti), 1,000 MPa (CoCr), and 270 MPa (PEKK). All loading simulations with the PEKK bar showed higher stresses in the implants and in the bone bed as well as higher displacements of the over denture in comparison to metal bars. PEKK showed different mechanical behaviour compared to Ti and CoCr. The distribution of stresses within the PEKK bar was wider than the area of loading which probably leads to fatigue of the whole bar and not only the part under load.
Reference32 articles.
1. Lambada, D, Lambada, P, Gundawar, S. Implant supported mandibular overdenture: a viable treatment option for edentulous mandible. J Clin Diagn Res 2014;8:ZD04–6.
2. Emami, E, de Souza, RF, Kabawat, M, Feine, JS. The impact of edentulism on oral and general health. Int J Dent 2013;2013:498305. https://doi.org/10.1155/2013/498305.
3. Heydecke, G, Thomason, JM, Lund, JP, Feine, JS. The impact of conventional and implant supported prostheses on social and sexual activities in edentulous adults: results from a randomized trial 2 months after treatment. J Dent 2005;33:649–57. https://doi.org/10.1016/j.jdent.2005.01.003.
4. Awad, MA, Lund, JP, Dufresne, E, Feine, JS. Comparing the efficacy of mandibular implantretained overdentures and conventional dentures among middle-aged edentulous patients: satisfaction and functional assessment. Int J Prosthodont (IJP) 2003;16:117–22.
5. Naert, I, Gizani, S, Vuylsteke, M, van Steenberghe, D. A 5-year prospective randomized clinical trial on the influence of splinted and unsplinted oral implants retaining a mandibular overdenture: prosthetic aspects and patient satisfaction. J Oral Rehabil 1999;26:195–202. https://doi.org/10.1046/j.1365-2842.1999.00369.x.
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献