External Quality Assessment in The Netherlands: time to introduce commutable survey specimens. Lessons from the Dutch “Calibration 2000” project

Author:

Baadenhuijsen Henk,Kuypers Aldy,Weykamp Cas,Cobbaert Christa,Jansen Rob

Abstract

AbstractThe performance of suitable secondary reference material for the use of trueness control of six routinely measured clinical enzymes in the Dutch External Quality Assessment (EQA) scheme is described. The reference material of choice was selected using the split-patient-sample between-field method (twin study) design as described in an earlier study of the Calibration 2000 project in The Netherlands. This material, which was proven to be commutable for all wet chemistry systems, was implemented as the national enzyme calibrator. It consisted of a cryo-protected lyophilised serum with additions of recombinant human enzymes. Various batches of the frozen version of this material without cryo-protection additive, called native EQA samples, were used in the general EQA scheme for performance evaluation. The results of Calibration 2000 calibrated and non-Calibration 2000 calibrated laboratories were compared for both the regular (spiked with non-human enzymes) and native EQA samples in terms of precision and bias with established reference method values for the native samples. The regular samples showed mean between-laboratory CV ranges for all six enzymes involved (low–high) of 5.5–10.3% for the non-calibrated users vs. 4.6–10.8% for the calibrated users. For the native samples these respective ranges were 5.2–9.9% vs. 2.2–4.9%. Without exception, the group of Calibration 2000 calibrated users showed the lowest bias against the reference method values. Regular EQA samples (spiked with non-human enzymes) showed poorer performance than native samples and are not suitable for accuracy assessment purposes, the main aim of EQA schemes. Native samples that are commutable should be used for trueness control in current EQA schemes.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Reference19 articles.

1. Multicenter evaluation of the commutability of a potential reference material for harmonization of enzyme activities;Scharnhorst;Clin Chem Lab Med,2004

2. Commutability assessment of potential reference material using a split - patient - sample between - field methods ( twin - study ) design : study within the framework of the Dutch project Calibration;Baadenhuijsen;Clin Chem,2000

3. Impact of reference materials on accuracy in clinical chemistry;Franzini;Clin Biochem,1998

4. Usefulness of reference materials in calibration of enzyme activities;Lessinger;Eur J Clin Chem Clin Biochem,1995

5. Trueness verification in European external quality assessment schemes : time to care about the quality of samples;Thienpont;Scand Clin Lab Invest,2003

Cited by 30 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. External quality assurance in the era of standardization;Clinica Chimica Acta;2024-04

2. Redesigning the surveillance of in vitro diagnostic medical devices and of medical laboratory performance by quality control in the traceability era;Clinical Chemistry and Laboratory Medicine (CCLM);2022-12-22

3. External quality control in laboratory medicine. Progresses and future;Advances in Laboratory Medicine / Avances en Medicina de Laboratorio;2022-08-17

4. Control externo de la calidad en medicina del laboratorio. Avances y futuro;Advances in Laboratory Medicine / Avances en Medicina de Laboratorio;2022-08-12

5. Control interno de la calidad – bases del pasado, situación presente y futuras tendencias;Advances in Laboratory Medicine / Avances en Medicina de Laboratorio;2022-05-26

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3