Multicentre performance evaluation of the E170 Module for MODULAR ANALYTICS

Author:

Bieglmayer Christian,Chan Daniel W.,Sokoll Lori,Imdahl Roland,Kobayashi Masaji,Yamada Erike,Lilje Diana J.,Luthe Hilmar,Meissner Jochen,Messeri Gianni,Celli Alessandra,Tozzi Paola,Roth Heinz-Jürgen,Schmidt Frank-Peter,Mächler Marie-Luise,Schuff-Werner Peter,Zingler Christiana,Smitz Johan,Schiettecatte Johan,Vonderschmitt Dieter J.,Pei Patrick,Ng Katherine,Ebert Christoph,Kirch Peter,Wanger Michael,McGovern Margaret,Stockmann Wolfgang,Kunst Albert

Abstract

AbstractThe E170 module was evaluated at 13 sites in an international multicentre study. The objective of the study was to assess the analytical performance of 49 analytes, and to collect feedback on the system’s reliability and practicability. The typical, within-run coefficients of variation (CVs) for most of the quantitative assays ranged between 1 and 2% while a range of 2–4% was achieved with the infectious disease methods. Total precision CVs were found to be within the manufacturer’s expected performance ranges, demonstrating good concordance of the system’s measuring channels and a high reproducibility during the 2–4-week trial period. The functional sensitivity of 11 selected assays met the clinical requirements (e.g., thyreotroponin (TSH) 0.008 mU/l, troponin T 0.02 µg/l, total prostate-specific antigen (PSA) 0.03 µg/l). The E170 showed no drift during an 8-hour period and no relevant reagent carryover. Accuracy was confirmed by ring trial experiments and method comparisons vs. Elecsys

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

Reference8 articles.

1. An alternative approach to a system evaluation in the field abstract;Bablok;Quim Clin,1995

2. Current databases on biological variation : pros cons and progress;Ricos;Scand Clin Lab Invest,1999

3. Myocardial infarction redefined : role of cardiac troponin testing;Apple;Clin Chem,2001

4. CAEv a program for computer aided evaluation;Bablok;J Autom Chem,1991

5. Assessment of the Roche / Hitachi MODULAR ANALYTICS under simulated routine conditions in European and US laboratories abstract;Stockmann;Clin Chem Lab Med,1999

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献