Author:
Moza Ajay,Gesenhues Jonas,Autschbach Rüdiger,Abel Dirk,Rossaint Rolf,Schmitz-Rode Thomas,Goetzenich Andreas
Abstract
AbstractBackground:Left ventricular assist device (LVAD) therapy has revolutionized the way end stage heart failure is treated today. Analysis of LVAD interaction with the whole cardiovascular system and its biological feedback loops is often conducted by means of computer models. Generating real time pressure volume loops (PV-loops) in patients, not using conductance catheters but routine diagnostics to feed anMethods:Routinely obtained hemodynamic measurements to evaluate myocardial function prior to LVAD implantation like pressure readings in the aorta, the left atrium and the left ventricle and simultaneous three-dimensional (3D) echocardiography recordings were assessed to parametrize a reduced computational model of the cardiovascular system. An automatic parameter identification procedure has been developed.Results:The results constitute a patient-individual computational simulation model. An exemplaryConclusion:The model improves understanding and prediction of the interaction between pump and ventricles. Future modifications in exporting and merging routinely assessed real time hemodynamic patient data are necessary to investigate various clinical and pathological conditions of LVAD recipients.
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献