Toth Arpad,Hasan Istabrak,Bourauel Christoph,Mundt Torsten,Biffar Reiner,Heinemann Friedhelm
Abstract
AbstractMini dental implants (MDI) were once thought of as transitional implants for treatment in selected clinical situations. Their reduced diameter makes them a very attractive option for patients with poor tolerance to maxillary and mandibular prostheses. Using the method of finite element analysis, a series of different designed MDI prototypes have been investigated. The prototypes differed in the geometry of implant body and/or design of implant head. The load transfer of the implant prototypes to the idealised alveolar bone has been regarded and the prototypes have been compared to each other and to a number of standard commercial implants. The prototype models have been virtually placed in the idealised bone with a cortical thickness of 1.5 mm and loaded laterally 30° from the implant's long axis. The condition of immediate loading was assumed for the numerical analyses through defining a contact interface between the implant and bone bed. The numerical analysis in this study showed that the design of the investigated prototype MDI of group 3 (mini-ball head) is the most advantageous design.