Author:
Hoffmann Nico,Weidner Florian,Urban Peter,Meyer Tobias,Schnabel Christian,Radev Yordan,Schackert Gabriele,Petersohn Uwe,Koch Edmund,Gumhold Stefan,Steiner Gerald,Kirsch Matthias
Abstract
AbstractMultimodal medical image fusion combines information of one or more images in order to improve the diagnostic value. While previous applications mainly focus on merging images from computed tomography, magnetic resonance imaging (MRI), ultrasonic and single-photon emission computed tomography, we propose a novel approach for the registration and fusion of preoperative 3D MRI with intraoperative 2D infrared thermography. Image-guided neurosurgeries are based on neuronavigation systems, which further allow us track the position and orientation of arbitrary cameras. Hereby, we are able to relate the 2D coordinate system of the infrared camera with the 3D MRI coordinate system. The registered image data are now combined by calibration-based image fusion in order to map our intraoperative 2D thermographic images onto the respective brain surface recovered from preoperative MRI. In extensive accuracy measurements, we found that the proposed framework achieves a mean accuracy of 2.46 mm.
Reference50 articles.
1. Survey of medical image registration;J Biomed Eng Tech,2013
2. Fusion of geometric and thermographic data for automated defect detection;J Electron Imaging,2012
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献