Excess of radiation burden for young testicular cancer patients using automatic exposure control and contrast agent on whole-body computed tomography imaging

Author:

Niiniviita Hannele12,Kulmala Jarmo13,Pölönen Tuukka4,Määttänen Heli1,Järvinen Hannu5,Salminen Eeva35

Affiliation:

1. Department of Medical Physics , Turku University Hospital , Turku , Finland

2. Department of Diagnostic Radiology , University of Turku , Turku , Finland

3. Oncology and Radiotherapy , Turku University Hospital , Turku , Finland

4. Department of Biostatistics , University of Turku , Turku , Finland

5. STUK-Radiation and Nuclear Safety Authority , Helsinki , Finland

Abstract

Abstract Background The aim of the study was to assess patient dose from whole-body computed tomography (CT) in association with patient size, automatic exposure control (AEC) and intravenous (IV) contrast agent. Patients and methods Sixty-five testicular cancer patients (mean age 28 years) underwent altogether 279 whole-body CT scans from April 2000 to April 2011. The mean number of repeated examinations was 4.3. The GE LightSpeed 16 equipped with AEC and the Siemens Plus 4 CT scanners were used for imaging. Whole-body scans were performed with (216) and without (63) IV contrast. The ImPACT software was used to determine the effective and organ doses. Results Patient doses were independent (p < 0.41) of patient size when the Plus 4 device (mean 7.4 mSv, SD 1.7 mSv) was used, but with the LightSpeed 16 AEC device, the dose (mean 14 mSv, SD 4.6 mSv) increased significantly (p < 0.001) with waist cirfumference. Imaging with the IV contrast agent caused significantly higher (13% Plus 4, 35% LightSpeed 16) exposure than non-contrast imaging (p < 0.001). Conclusions Great caution on the use of IV contrast agent and careful set-up of the AEC modulation parameters is recommended to avoid excessive radiation exposure on the whole-body CT imaging of young patients.

Publisher

Walter de Gruyter GmbH

Subject

Radiology, Nuclear Medicine and imaging,Oncology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3