Objectives
Cellular analysis of body fluids (BFs) can assist clinicians for the diagnosis of many medical conditions. The aim of this work is the evaluation of the analytical performance of the UF-5000 body fluid mode (UF-BF) analyzer compared to the gold standard method (optical microscopy, OM) and to XN-1000 (XN-BF), another analyzer produced by the same manufacturer (Sysmex) and with a similar technology for BF analysis.
Methods
One hundred BF samples collected in K3EDTA tubes were analyzed by UF-BF, XN-BF and OM. The agreement was evaluated using Passing and Bablok regression and Bland–Altman plot analysis. The receiver operating characteristic (ROC) curves were selected for evaluating the diagnostic agreement between OM classification and UF-BF parameters.
Results
Comparison between UF-BF and OM, in all BF types, showed Passing and Bablok’s slope comprised between 0.99 (polymorphonuclear cells count, PMN-BF) and 1.39 (mononuclear cells count, MN-BF), the intercepts ranged between 26.47 (PMN-BF parameter) and 226.80 (white blood cell count). Bland–Altman bias was comprised between 7.3% (total cell count, TC-BF) and 52.9% (MN-BF). Comparison between UF-BF and XN-BF in all BF showed slopes ranged between 1.07 (TC-BF and PMN-BF) and 1.16 (MN-BF), intercepts ranged between 8.30 (PMN) and 64.78 (WBC-BF). Bland–Altman bias ranged between 5.8 (TC-BF) and 21.1% (MN-BF). The ROC curve analysis showed an area under the curve ranged between 0.9664 and 1.000.
Conclusions
UF-BF shows very good performance for the differential counts of cells in ascitic, pleural and synovial fluids and therefore it is useful to screen and count cells in this type of BF.
Subject
Biochemistry (medical),Clinical Biochemistry,Public Health, Environmental and Occupational Health,Health Policy,Medicine (miscellaneous)
Reference29 articles.
1. Sandhaus, LM. Body fluid cell counts by automated methods. Clin Lab Med 2015;35:93–103. https://doi.org/10.1016/j.cll.2014.10.003.
2. Fleming, C, Russcher, H, Lindemans, J, de Jonge, R. Clinical relevance and contemporary methods for counting blood cells in body fluids suspected of inflammatory disease. Clin Chem Lab Med 2015;53:1689–706. https://doi.org/10.1515/cclm-2014-1247.
3. Bignardi, GE. Flow cytometry for the microscopy of body fluids in patients with suspected infection. J Clin Pathol 2015;68:870–8. https://doi.org/10.1136/jclinpath-2015-203088.
4. Yang, D, Zhou, Y, Chen, B. Performance evaluation and result comparison of the automated hematology analyzers Abbott CD 3700, Sysmex XE 2100 and Coulter LH 750 for cell counts in serous fluids. Clin Chim Acta 2013;419:113–8. https://doi.org/10.1016/j.cca.2013.01.022.
5. Clinical Laboratory Standards Institute. Body fluid analysis for cellular composition; approved guideline. CLSI document H56-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2006.