Author:
Lopez-Calle Eloisa,Espindola Pamela,Spinke Juergen,Lutz Sascha,Nichtl Alfons,Tgetgel Armin,Herbert Nicolas,Marcinowski Moritz,Klepp Juergen,Fischer Thomas,Brueckner Thorsten,Boehm Christoph,Keller Thomas
Abstract
AbstractBackground:A multitude of troponin assays for the point-of-care (POC) have been developed showing a lack of analytical sensitivity and precision. We present a new platform solution for the high-sensitivity detection of cardiac troponin T (cTnT) in a 30 μL whole blood sample with a turnaround time of 11 min.Methods:The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents. After the sample application, the assay was automatically processed by separating the cellular blood components via centrifugation, followed by incubation of a defined volume from the generated plasma with the immunoreagents. The fluorescence in the signal zone of a membrane was measured after its washing for the cTnT quantitation.Results:A calibration curve, measured in whole blood samples spiked with native human cTnT, was generated covering a range up to a concentration of approximately 8300 ng/L. The lower detection limit was determined to be 3.0 ng/L. At a concentration of 14 ng/L, the 99th percentile value from the high-sensitivity cardiac troponin T (hs-cTnT) assay in the ElecsysConclusions:The described technology shows that an analytical performance for a highly sensitive determination of cTnT can be achieved in a POC setting.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Cited by
11 articles.