Auto-validation of complete blood counts in an outpatient’s regional laboratory

Author:

Froom Paul,Barak Mira

Abstract

AbstractThe rate of auto-validation is dependent on the ability of the laboratory information system (LIS) to integrate historical data, on the frequency and methods for identifying analyzer errors, and on the criteria for reflex testing, including the need for peripheral smear review. The rate of auto-validation in outpatient laboratories, however, is unclear.We examined 45,925 consecutive complete blood count (CBC) test results (1 January, 2014–31 January, 2014) from patients aged 50±24 years. The LIS auto-validates all samples according to set criteria. Technicians validated test results when previous CBC test results were required to determine: 1) the need for peripheral slide review and/or sample rerun or 2) the need for reflex testing to detect autoimmune hemolytic anemia or β-thalassemia minor.The auto-validation rates were 97.6% after rejecting results requiring validation to determine the need for a peripheral smear review and/or sample rerun. This decreased to 92.9% after including reflex testing to determine the reasons for normocytic and microcytic anemia. We estimated that auto-validation decreased the workload by 7.7–11.6 h per 3000 test results.We conclude that very high auto-validation rates are possible in outpatient general laboratories, leading to conformity in the validation process and a considerable estimated savings in technician time. Further studies are needed in other settings.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Reference36 articles.

1. An audit of the indications for the reporting of blood films : results from the National Study;Galloway;Pathology J Clin Pathol,2006

2. new and effective way for preventing pre - analytical laboratory errors;Barak;Clin Chem Lab Med,2014

3. for the Hematology And Clinical Microscopy Resource Committee of the College of American Pathologists Physician review of the peripheral blood smear : when and why an opinion;Peterson;Lab,2001

4. International Consensus Group for for Laboratory Hematology Suggested criteria for action following automated CBC WBC differential analysis Available from http www islh org Committees ConsensusGroup CGICGHREview htm Accessed on;March;Hematology Review International Society,2004

5. multicenter case - control study on predictive factors distinguishing childhood leukemia from juvenile rheumatoid arthritis;Jones;Pediatrics,2006

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献