Author:
Oris Charlotte,Chabanne Russel,Durif Julie,Kahouadji Samy,Brailova Marina,Sapin Vincent,Bouvier Damien
Abstract
Abstract
Background
The addition of S100B protein to guidelines for the management of mild traumatic brain injury (mTBI) decreases the amount of unnecessary computed tomography (CT) scans with a significant decrease in radiation exposure and an increase in cost savings. Both DiaSorin and Roche Diagnostics have developed automated assays for S100B determination. Recently, bioMérieux developed a prototype immunoassay for serum S100B determination. For the first time, we present the evaluation of the S100B measurement using a bioMérieux Vidas® 3 analyzer.
Methods
We evaluated the matrix effects of serum and plasma, and their stability after storage at 2–8 °C, −20 °C and −80 °C. The new measurement prototype (bioMérieux) was compared with an established one (Roche Diagnostics), and a precision study was also conducted. Lastly, clinical diagnostics performance of the bioMérieux and Roche Diagnostics methods were compared for 80 patients referred to the Emergency Department for mTBI.
Results
Stability after storage at 2–8 °C, −20 °C, and −80 °C and validation of the serum matrix were demonstrated. The bioMérieux analyzer was compared to the Roche Diagnostics system, and the analytical precision was found to be efficient. Clinical diagnosis performance evaluation confirmed the predictive negative value of S100B in the management of mTBI.
Conclusions
The study’s data are useful for interpreting serum S100B results on a bioMérieux Vidas® 3 analyzer.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献