Assignment of international normalized ratio to frozen and freeze-dried pooled plasmas

Author:

van den Besselaar Antonius M.H.P.12,Cobbaert Christa M.3

Affiliation:

1. Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine , Leiden University Medical Center , P.O. Box 9600 , 2300 RC Leiden , The Netherlands

2. Department of Thrombosis and Hemostasis , Leiden University Medical Center , Leiden , The Netherlands , Phone: +31 71 526 1942

3. Coagulation Reference Laboratory, Department of Clinical Chemistry and Laboratory Medicine , Leiden University Medical Center , Leiden , The Netherlands

Abstract

Abstract Objectives Frozen and freeze-dried plasmas may be used for local prothrombin time system calibration, for direct international normalized ratio (INR) determination, and for quality assessment. The purpose of the present study was to evaluate the usefulness of INRs assigned with various types of thromboplastins to frozen and freeze-dried pooled plasmas obtained from patients treated with vitamin K antagonists. Methods INRs were calculated according to the international sensitivity index (ISI) model using various thromboplastins and instruments, i.e. International Standards for thromboplastin as well as six commercial reagents prepared from rabbit and bovine brain, and recombinant human tissue factor. The uncertainty of the INRs was assessed using the standard deviations of clotting times and ISI values. Commutability of the plasmas was assessed according to the approved Clinical and Laboratory Standards Institute (CLSI) Guideline EP30-A. Validation of a set of six frozen plasma pools for direct INR determination was performed according to the Subcommittee on Control of Anticoagulation of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis (SSC/ISTH) guidelines. Results For all frozen and freeze-dried plasmas, the INRs calculated with bovine thromboplastin Thrombotest were lower than the INRs assigned with other thromboplastins. With a few exceptions, the frozen and freeze-dried pooled plasmas were commutable. When the set of six frozen plasma pools was used for local calibration, the analytical bias of the INR was less than ±10% for all commercial reagents except Thrombotest. Conclusions Processing of fresh plasmas to prepare pooled frozen plasmas and freeze-dried plasmas may lead to different INR assignments depending on the thromboplastin used. Despite minor INR differences, a set of six frozen plasma pools could be used for local calibration by direct INR determination.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine