Enhanced specificity due to method specific limits for relative ion intensities in a high-performance liquid chromatography – tandem mass spectrometry method for iohexol in human serum

Author:

Schweinsberg Vincent12,Eckelt Stefan2,Schulz Anna Marta3,Arlt Marcel2,Schumann Marc2,van der Giet Markus3,Tauber Rudolf12,Binscheck-Domaß Torsten2,Kießling Peggy12

Affiliation:

1. Charité – Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Institute for Laboratory Medicine, Clinical Chemistry and Pathobiochemistry, Berlin, Germany

2. Labor Berlin-Charité Vivantes GmbH, Department of Laboratory Medicine and Toxicology, Sylter Str. 2, 13353 Berlin, Germany

3. Charité – Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Department of Nephrology, Berlin, Germany

Abstract

AbstractBackgroundAccurate assessment of kidney function is needed for a variety of clinical indications and for research. The measurement of the serum clearance of iohexol has emerged as a feasible method to reach this objective. We report the analytical validation and clinical application of a new high-performance liquid chromatography (HPLC) – tandem mass spectrometry (MS/MS) assay to quantify iohexol in human serum. Specificity was enhanced due to the use of method specific acceptance limits for relative ion (RI) intensities.MethodsThe internal standard ioversol was added to 50 μL serum prior to protein precipitation with methanol. Linear gradient elution was performed on a Waters Oasis® HLB column. Three transitions for both iohexol and ioversol were monitored allowing calculation of RIs. Measurements acquired during method validation were used as a training set to establish stricter acceptance criteria for RIs which were then tested retrospectively on clinical routine measurements (86 measurements) and on mathematically simulated interferences.ResultsThe method was linear between 5.0 μg/mL (lower limit of quantification [LLOQ]) and 100.3 μg/mL iohexol. Intraday and interday imprecision were ≤2.6% and ≤3.2%, respectively. Bias was −1.6% to 1.5%. All validation criteria were met, including selectivity, recovery, extraction efficiency and matrix effects. Retrospectively acceptance limits for RIs could be narrowed to ±4 relative standard deviations of the corresponding RIs in the training set. The new limits resulted in an enhanced sensitivity for the simulated interferences.ConclusionsCriteria for validation were met and the assay is now used in our clinical routine diagnostics and in research.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Advancements in mass spectrometry as a tool for clinical analysis: Part I;Clinical Chemistry and Laboratory Medicine (CCLM);2020-04-28

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3