Measurement of sirolimus concentrations in human blood using an automated electrochemiluminescence immunoassay (ECLIA): a multicenter evaluation

Author:

Stove Veronique,Ramos Pedro Alía,Wallemacq Pierre,Vogeser Michael,Schuetzenmeister Andre,Schmiedel Christian,Shipkova Maria

Abstract

AbstractBackground:Therapeutic drug monitoring (TDM) of sirolimus is essential in transplant recipients. We evaluated the performance of a new electrochemiluminescence immunoassay (ECLIA) for measuring sirolimus concentrations in whole blood at five European laboratories.Methods:Study assessments included repeatability, intermediate precision and functional sensitivity (concentration at coefficient of variation [CV] of 20%) experiments. Method comparisons with liquid chromatography-tandem mass spectrometry (LC-MS/MS; reference method) and two immunoassays (chemiluminescent microparticle immunoassay [CMIA] and antibody-conjugated magnetic immunoassay [ACMIA]) were performed using native samples from patients with kidney transplants.Results:Imprecision testing CVs were ≤6.4% and ≤10.7% across the sirolimus concentration range for both repeatability and intermediate precision, respectively. The ECLIA showed excellent functional sensitivity: the CV did not reach 20%; the CV at the assay’s limit of quantitation (1.5 μg/L) was 7.0%. Agreement between the ECLIA and LC-MS/MS using native kidney samples was close, with weighted Deming regression analysis yielding a slope of 1.05, an intercept of 0.154 μg/L and a Pearson’s correlation coefficient (r) of 0.94, while Bland-Altman analysis showed a combined mean bias of 0.41 μg/L (±2 standard deviation [SD], −1.96 to 2.68). The ECLIA also showed good correlation with the two other immunoassays: the CMIA (slope=0.91, intercept=0.112 μg/L and r=0.89) and the ACMIA (slope=0.99, intercept=0.319 μg/L and r=0.97).Conclusions:The ECLIA showed good precision, functional sensitivity and agreement with other methods of sirolimus measurement used in clinical practice, suggesting that the assay is suitable for TDM in transplant recipients and provides an alternative to LC-MS/MS.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Reference34 articles.

1. Statistical methods for assessing agreement between two methods of clinical measurement;Lancet,1986

2. Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?;Clin Chem,2012

3. Limit of blank, limit of detection and limit of quantitation;Clin Biochem Rev,2008

4. Quantification of sirolimus by liquid chromatography-tandem mass spectrometry using on-line solid-phase extraction;Clin Chem Lab Med,2002

5. Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?;Clin Chem,2012

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3