Stove Veronique,Ramos Pedro Alía,Wallemacq Pierre,Vogeser Michael,Schuetzenmeister Andre,Schmiedel Christian,Shipkova Maria
Abstract
AbstractBackground:Therapeutic drug monitoring (TDM) of sirolimus is essential in transplant recipients. We evaluated the performance of a new electrochemiluminescence immunoassay (ECLIA) for measuring sirolimus concentrations in whole blood at five European laboratories.Methods:Study assessments included repeatability, intermediate precision and functional sensitivity (concentration at coefficient of variation [CV] of 20%) experiments. Method comparisons with liquid chromatography-tandem mass spectrometry (LC-MS/MS; reference method) and two immunoassays (chemiluminescent microparticle immunoassay [CMIA] and antibody-conjugated magnetic immunoassay [ACMIA]) were performed using native samples from patients with kidney transplants.Results:Imprecision testing CVs were ≤6.4% and ≤10.7% across the sirolimus concentration range for both repeatability and intermediate precision, respectively. The ECLIA showed excellent functional sensitivity: the CV did not reach 20%; the CV at the assay’s limit of quantitation (1.5 μg/L) was 7.0%. Agreement between the ECLIA and LC-MS/MS using native kidney samples was close, with weighted Deming regression analysis yielding a slope of 1.05, an intercept of 0.154 μg/L and a Pearson’s correlation coefficient (r) of 0.94, while Bland-Altman analysis showed a combined mean bias of 0.41 μg/L (±2 standard deviation [SD], −1.96 to 2.68). The ECLIA also showed good correlation with the two other immunoassays: the CMIA (slope=0.91, intercept=0.112 μg/L and r=0.89) and the ACMIA (slope=0.99, intercept=0.319 μg/L and r=0.97).Conclusions:The ECLIA showed good precision, functional sensitivity and agreement with other methods of sirolimus measurement used in clinical practice, suggesting that the assay is suitable for TDM in transplant recipients and provides an alternative to LC-MS/MS.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference34 articles.
1. Statistical methods for assessing agreement between two methods of clinical measurement;Lancet,1986
2. Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?;Clin Chem,2012
3. Limit of blank, limit of detection and limit of quantitation;Clin Biochem Rev,2008
4. Quantification of sirolimus by liquid chromatography-tandem mass spectrometry using on-line solid-phase extraction;Clin Chem Lab Med,2002
5. Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?;Clin Chem,2012
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献