Abstract
Abstract
Background
Plasma renin levels were determined in the academia-driven, EU-funded “Labeling of Enalapril from Neonates up to Adolescents” (LENA) project to evaluate its role in pediatric heart failure. Quality-controlled bioanalysis is crucial to ensure reliable data generation. However, a comprehensive bioanalytical quality control (QC) concept to monitor the method performance within an academic environment was lacking.
Methods
Thus, a QC concept was designed encompassing regulatory guidance, international recommendations and current scientific discussions. The concept included (1) a system-suitability test, (2) verification of single bioanalytical runs by calibration curve performance and evaluation of QCs, (3) assessment of the inter-run accuracy according to Clinical Laboratory Standards Institute (CLSI) guideline, (4) monitoring of reproducibility by pediatric incurred samples, (5) blank-sample analysis and (6) participation in interlaboratory testing.
Results
The concept was successfully applied to the academic project. About 11% of single runs were identified as invalid and triggered a re-analysis of unknown samples being included in those runs. The usefulness of the customized inter-run monitoring was demonstrated and proved the good accuracy from the first to the last run. All 147 reanalyzed incurred sample pairs complied with regulatory requirements.
Conclusions
The regulatory complied QC concept was customized for the demands of academia-driven pediatric trials and contributed to the reliable quantification of 965 pediatric renin samples.
Funder
LENA collaborator list – (‘on behalf of the LENA consortium’): LENA Beneficiaries and Principal Investigators
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献