Development and validation of LC-MS/MS methods to measure tobramycin and lincomycin in plasma, microdialysis fluid and urine: application to a pilot pharmacokinetic research study

Author:

Pandey Saurabh1,Dhanani Jayesh23,Lipman Jeffrey23,Roberts Jason A.2345,Wallis Steve C.2,Parker Suzanne L.2

Affiliation:

1. UQ Centre for Clinical Research, Faculty of Medicine , The University of Queensland , Brisbane , QLD 4029 , Australia , Phone: +61 7 33465104

2. UQ Centre for Clinical Research, Faculty of Medicine , The University of Queensland , Brisbane , Australia

3. Department of Intensive Care Medicine , Royal Brisbane and Women’s Hospital , Brisbane , Australia

4. Department of Pharmacy , Royal Brisbane and Women’s Hospital , Brisbane , Australia

5. Centre of Translational Anti-Infective Pharmacodynamics, School of Pharmacy , The University of Queensland , Brisbane , Australia

Abstract

Abstract Background The aim of our work was to develop and validate a hydrophilic interaction liquid chromatography-electrospray ionization-tandem mass spectrometry (HILIC-ESI-MS/MS) methods for the quantification of tobramycin (TMC) and lincomycin (LMC)in plasma, microdialysis fluid and urine. Methods Protein precipitation was used to extract TMC and LMC from plasma, while microdialysis fluid and urine sample were diluted prior to instrumental analysis. Mobile phase A consisted of 2 mM ammonium acetate in 10% acetonitrile with 0.2% formic acid (v/v) and mobile phase B consisted of 2 mM ammonium acetate in 90% acetonitrile with 0.2% formic acid (v/v). Gradient separation (80%–10% of mobile phase B) for TMC was done using a SeQuant zic-HILIC analytical guard column. While separation of LMC was performed using gradient elution (100%–40% of mobile phase B) on a SeQuant zic-HILIC analytical column equipped with a SeQuant zic-HILIC guard column. Vancomycin (VCM) was used as an internal standard. A quadratic calibration was obtained over the concentration range for plasma of 0.1–20 mg/L for TMC and 0.05–20 mg/L for LMC, for microdialysis fluid of 0.1–20 mg/L for both TMC and LMC, and 1–100 mg/L for urine for both TMC and LMC. Results For TMS and LMC, validation testing for matrix effects, precision and accuracy, specificity and stability were all within acceptance criteria of ±15%. Conclusions The methods described here meet validation acceptance criteria and were suitable for application in a pilot pharmacokinetic research study performed in a sheep model.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3