Comparison of three different chemiluminescence assays and a rapid liquid chromatography tandem mass spectrometry method for measuring serum aldosterone

Author:

Yin Yicong1,Ma Chaochao1,Yu Songlin1,Liu Wenjing2,Wang Danchen1,You Tingting1,Cheng Qian1,Qiu Ling1

Affiliation:

1. Department of Clinical Laboratory , Peking Union Medical College Hospital, Peking Union Medical College and Chinese Academy of Medical Science , Beijing , P.R. China

2. Waters Corporation Shanghai Science and Technology Co. Ltd. , Beijing , P.R. China

Abstract

Abstract Background This study aimed to quantify and compare serum aldosterone (sALD) levels through three different chemiluminescence immunoassays (CLIAs) and liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis. Methods Serum samples from 221 patients with suspected primary aldosteronism (PA) were retrospectively included in this study conducted at the Peking Union Medical College Hospital from June to August in 2017. sALD levels were determined using the LC-MS/MS method and three different CLIA systems, viz., DiaSorin® XL, iSYS and Auto Lumo A2000. Pooled fresh serum samples were used for recalibration. Passing-Bablok regression analysis, correlation matrix, and Bland-Altman plots were used to evaluate the concurrence among ALD levels determined using the three CLIAs. Results Within-laboratory precision of the four assays ranged from 2.1% to 9.4%, except the coefficient variation (CV) of one of the CLIAs, which exceeded 20.0% for samples with low sALD levels. sALD levels determined using LC-MS/MS were significantly lower than those determined using the other three CLIAs (p < 0.0001). Spearman’s correlation coefficient of the four assays ranged from 0.745 to 0.950 (p < 0.0001). The Bland-Altman plot showed that the average bias (%) for the three CLIAs and LC-MS/MS ranged from −69.3 to −49.2. After recalibration, this correlation did not improve among the assays. However, the bias and bias percentage at the medical decision level improved between LC-MS/MS and DiaSorin® XL/iSYS. Conclusions Significant inconsistencies between the results of CLIAs and LC-MS/MS indicate that different sALD measures cannot be used interchangeably.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3