Integrity of serum samples is changed by modified centrifugation conditions

Author:

Miler Marijana1,Nikolac Gabaj Nora2,Culej Jelena2,Unic Adriana2,Vrtaric Alen2,Milevoj Kopcinovic Lara2

Affiliation:

1. Department of Clinical Chemistry , Sestre Milosrdnice University Hospital Center , Vinogradska 29 , Zagreb , Croatia

2. Department of Clinical Chemistry , Sestre Milosrdnice University Hospital Center , Zagreb , Croatia

Abstract

Abstract Background Serum samples should be centrifuged for at least 10 min at 1300–2500 × g. Changed centrifugation conditions could compromise sample quality. The objective of this study was to compare the serum quality and turnaround time (TAT) using different centrifugation conditions. Methods The study was done in four different periods (A, B, C and D) at different conditions: for 10, 5 and 7 (A, B and C, respectively) at 2876 × g, and 7 (D) min at 4141 × g. Sample quality was assessed as the proportion of samples with: (a) aspiration errors, (b) H index >0.5 g/L and (c) suppressed reports of potassium (K) due to hemolysis. TAT was calculated for emergency samples. The proportions of samples (a), (b) and (c) were compared according to period A. Results The number of aspiration errors was significantly higher in samples centrifuged at 2876 × g for 5 min (p = 0.021) and remained unchanged when centrifuged for 7 min (p = 0.066 and 0.177, for periods C and D, respectively). In periods B, C and D, the proportion of samples with hemolysis was higher than that in period A (p-values 0.039, 0.009 and 0.042, respectively). TAT differed between all periods (p < 0.001), with the lowest TAT observed for B and D. The lowest number of samples exceeding 60-min TAT was observed in period D (p = 0.011). Conclusions The integrity of serum samples is changed with different centrifugation conditions than those recommended. Our study showed that shorter centrifugation at higher force (7 min at 4141 × g) significantly decreases TAT, with unchanged proportion of samples with aspiration errors.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry, medical,Clinical Biochemistry,General Medicine

Reference29 articles.

1. Clinical Laboratory Standards Institute. Procedures for the Handling and Processing of Blood Specimens for Common Laboratory Tests; Approved Guideline – Third Edition. Wayne, PA: Clinical and Laboratory Standards Institute, 2004. CLSI document H18-A3.

2. Greiner Bio-One. Vacuette blood collection system – handling recommendations. 980183 rev 05, February 2016. http://www.gbo.com/preanalytics. Accessed: 25 Feb 2019.

3. Becton Dickinson. BD Diagnostics – Preanalytical Systems. 2009. https://www.bd.com/resource.aspx?IDX=10155. Accessed: 25 Feb 2019.

4. World Health Organization. Use of anticoagulants in diagnostic laboratory investigations, 2nd ed. Geneva: World Health Organization, 2002.

5. Holland LL, DomBourian MD. Evaluation of an abbreviated centrifugation protocol for chemistry testing. Lab Med 2012;43:78–81.