Abstract
Abstract
Background
Development of automated analyzers for erythrocyte sedimentation rate (ESR) has imposed the need for extensive validation prior to their implementation in routine practice, to ensure comparability with the reference Westergren method. The aim of our study was to perform the analytical validation of two automated ESR analyzers, the Ves-Matic Cube 200 and the TEST1.
Methods
Validation was performed according to the recent International Council for Standardization in Hematology recommendations and included determination of intrarun and inter-run precision, assessment of sample carryover, hemolysis interference, sensitivity to fibrinogen, method comparison with the gold standard Westergren method and stability test.
Results
The highest intrarun imprecision was obtained for the low ESR range (33.5% for Ves-Matic Cube; 37.3% for TEST1) while inter-run coefficients of variation on three levels were much better for the TEST1 (0%, 2% and 1.2%) compared to the Ves-Matic Cube 200 on two levels (24.9% and 5.8%). Both Ves-Matic Cube 200 and TEST1 showed no statistically significant difference when compared with Westergren. Bland-Altman analysis yielded overall insignificant mean biases for all comparisons, but a wider dispersion of results and 95% limits of agreement for comparisons including the Ves-Matic Cube 200. Carryover was considered insignificant, while hemolysis had a negative effect on all assessed ESR methods. The highest sensitivity to fibrinogen was observed for the Ves-Matic Cube 200, followed by Westergren and the least sensitive was the TEST1.
Conclusions
The obtained results proved the analytical validity of the TEST1 and the Ves-Matic Cube 200, and high comparability with the gold standard Westergren method, showing obvious improvements in standardization of ESR methods.
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference56 articles.
1. Evaluation of the TEST 1 erythrocyte sedimentation rate system and intra- and inter-laboratory quality control using new latex control materials;Clin Chem Lab Med,2010
2. International Council for Standardization in Haematology. ICSH review of the measurement of the erythrocyte sedimentation rate;Int J Lab Hematol,2011
3. Erythrocyte sedimentation rate: innovative techniques for an obsolete test?;Clin Chem Lab Med,2003
4. The plasma protein fibrinogen stabilizes clusters of red blood cells in microcapillary flows;Sci Rep,2014
5. General assembly, diagnosis, laboratory test: proceedings of International Orthopedic Infections;J Arthoplasty,2019
Cited by
28 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献