Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay

Author:

Lee Jehoon1,Kim Myungshin1,Chae Hyojin1,Kim Yeongsic1,Park Hae-il1,Kim Yonggoo1,Chi Hyunyoung2,Kwon Hi Jeong1

Affiliation:

1. Department of Laboratory Medicine, College of Medicine, The Catholic University of Korea , Seoul , Korea

2. Samkwang Medical Laboratories , Seoul , Korea

Abstract

Abstract Background: A novel enzymatic HbA1c assay was introduced for use in an automated chemistry analyzer. With this unique method, HbA1c and plasma glucose can be measured from the same EDTA tube. We evaluated the analytical performance of this enzymatic HbA1c assay in a JCA-BM6010/C analyzer and compared the HbA1c values with the results from other widely used methodological instruments. Methods: The imprecision, linearity, carry-over and concordance rate of the enzymatic HbA1c test (BM Test HbA1c) using the JCA-BM6010/C analyzer were evaluated. Three hundred and seventy-seven specimens with HbA1c concentrations from 16 to 133 mmol/mol were used for a comparison study with two high performance liquid chromatography methods: Variant II Turbo and Tosoh HLC 723 G8 and the AutoLab Hemoglobin A1c immunoturbidimetry reagent using a Hitachi 7600-110. Forty specimens were used for the glucose method comparison. Results: The HbA1c coefficients of variation of the within-run imprecision for low and high levels were 0.6% and 0.4%, respectively. The linearity of the BM Test HbA1c using the JCA-BM6010/C analyzer was excellent in the range between 31 mmol/mol and 143 mmol/mol. The carry-over rate was 0.2%. The relationships between the BM test and the other three methods were 0.916×Tosoh G8+3.644, r=0.986; 0.887×Bio-Rad Variant II+1.896, r=0.972; and 0.941×AutoLab+4.532, r=0.977. The concordance rates using a cut-off of 48 mmol/mol were 91.5% with Tosoh G8, 82.8% with Bio-Rad Variant II, and 91.0% with AutoLab. The simultaneously assayed plasma glucose with HbA1c was 1.002×Routine plasma glucose+0.625, r=1.000 Conclusions: The enzymatic BM Test HbA1c in the JCA-BM6010/C analyzer showed excellent precision and linearity, and a minimal carry-over rate. The simultaneously assayed plasma glucose analysis showed good performance.

Publisher

Walter de Gruyter GmbH

Subject

Biochemistry (medical),Clinical Biochemistry,General Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3