Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study

Author:

Schenk Martin1,Müller Gottfried2,Greiner Tim Oliver1,Fahrner Christine1,Königsrainer Alfred1,Thiel Christian1

Affiliation:

1. 1Universitätsklinikum Tübingen, Klinik für Allgemeine, Viszeral- und Transplantationschirurgie, Institut für Experimentelle Chirurgie, Tübingen, Germany

2. 2Caritas Krankenhaus, Klinik für Allgemeine, Visceral- und Gefäßchirurgie, Bad Mergentheim, Germany

Abstract

AbstractBackground: The use of surgical ligation clips is considered as the gold standard for the closure of vessels, particularly in laparoscopic surgery. The safety of clips is mainly achieved by the deep indentation of the metal bar with a high retention force. A novel double-shanked (DS) titanium clip was compared to two single-shanked clips with respect to axial and radial pull-off forces.Methods: In a porcine model (8 animals, 51±1 kg), arteries were prepared immediately after euthanisation, assigned to either a medium (2–4 mm; n=120) or a medium-large (3.5–7 mm; n=120) clip size group, and clipped with the appropriate clip size. After dissection, axial and radial pull-off forces were measured.Results: The axial pull-off force of the DS-Clip was higher than one single-shanked clip and comparable to the other single-shanked clip, and overall was linearly correlated to the cross-sectional area of the clip. The radial pull-off force of the DS-Clip was significantly higher than both single-shanked clips and, for the single-shanked clips, was correlated to the total clip thickness. The variation of radial pull-off force was lower for the DS-Clip due to a defined catch in the clip applier.Conclusions: The radial pull-off force was lower than the axial pull-off force in total and therefore appears to be the critical point of dislocation. Due to the higher total holding mass, the DS-Clip provided a clear advantage in this regard and might therefore decrease the dislocation rate. The catch in the applier increases the reproducibility in clip placement.

Publisher

Walter de Gruyter GmbH

Subject

Surgery

Reference30 articles.

1. Appendix stump closure with titanium clips in laparoscopic appendectomy;Langenbecks Arch Surg,2012

2. Bile Duct “clip-stones” – why a stitch in time could save nine;Hepatogastroenterology,2010

3. Laparoscopically placed clips on the human cystic duct – an experimental comparison of dislocation between resorbable and titanium clips;Chirurg,1997

4. Dislocation of clips of the cystic duct stump;Zentralbl Chir,1998

5. Comparison of various methods of vessel ligation: what is the safest method?;Surg Endosc,2013

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献