A Comparison of ETV and LA for the Determination of Trace Elements in Solid Samples by MIP OES

Author:

Matusiewicz Henryk1,Ślachciński Mariusz1

Affiliation:

1. Faculty of Chemical Technology , Poznan University of Technology , ul. Berdychowo 4, 60-965 Poznań , Poland , phone +48 61 665 23 12, fax +48 61 665 25 71

Abstract

Abstract The performance of electrothermal vaporization (ETV) and laser ablation (LA) of dry aerosols as sample introduction systems for microwave induced plasma optical emission spectrometry (MIP OES) are compared and evaluated in terms of detection limits, precision and accuracy for the determination of trace elements (Ca, Cd, Cu, Fe, Mg, Mn, Sr, Zn) in the same solid micro samples. In MIP OES both radiation sources can be independently adjusted to optimize the sampling process and then its subsequent excitation. A univariate approach and simplex optimization procedure were used to obtain the best signal/noise (S/N) ratio and derive analytical figures of merit. A comparison using a Student’s t-test between the results obtained by both ETV/LA-MIP OES methods for trace elements, and concentrations in standard reference material (SRM) and certified reference materials (CRMs) showed that there was no significant differences on a 95 % confidence level. The detection limits of the tested elements in solid samples by ETV/LA-MIP OES were in the range of 0.1 to 11 µg g−1 for all elements determined, while the corresponding absolute values in the range of ng. The precision of the results for ETV-MIP OES and LA-MIP OES varied between 2 and 4 % and 3 and 7 %, respectively. The linear dynamic ranges in the ETV/LA-MIP OES are extend over three decades of concentration. The methods were validated by the analysis of NIST SRM 2711Montana Soil, NRCC CRM PACS-2 Marine Sediment and NRCC CRM TORT-2 Lobster Hepatopancreas of different matrix composition and by the standard addition technique.

Publisher

Walter de Gruyter GmbH

Subject

Environmental Chemistry,Environmental Engineering

Reference10 articles.

1. [1] Tendero C, Tixier C, Tristant P, Desmaison J, Leprince P. Atmospheric pressure plasmas: A review. Spectrochim Acta B. 2006;61:2-30. DOI: 10.1016/j.sab.2005.10.003.10.1016/j.sab.2005.10.003

2. [2] Mester Z, Sturgeon R, editors. Sample Preparation for Trace Element Analysis. Amsterdam: Elsevier; 2003. ISBN 9781849730525.

3. [3] Jankowski KJ, Reszke E. Microwave Induced Plasma Analytical Spectrometry. Cambridge. UK: Royal Society of Chemistry; 2011. ISBN 9780444511010.

4. [4] Matusiewicz H. Thermal vaporization for sample introduction in microwave induced plasma analytical emission spectrometry. Spectrochim Acta Rev. 1990;13:47-68.

5. [5] Resano M, Vanhaecke F, de Loos-Vollebregt MTC. Electrothermal vaporization for sample introduction in atomic absorption, atomic emission and plasma mass spectrometry - a critical review with focus on solid sampling and slurry analysis. J Anal At Spectrom. 2008;23:1450-1475. DOI: 10.1039/b807756h.10.1039/b807756h