Sub-3mm spatial resolution from a large monolithic LaBr3 (Ce) scintillator

Author:

Liprandi Silvia1,Mayerhofer Michael23,Aldawood Saad45,Binder Tim2,Dedes George2,Miani Agnese26,Schaart Dennis R.7,Lozano Ingrid I. Valencia2,Parodi Katia2,Thirolf Peter G.2

Affiliation:

1. Ludwig Maximilians Universität (LMU) München, Medical Physics Department, Am Coulombwall 1, Garching, Germany

2. LMU München, Medical Physics Department, Germany

3. University of Hamburg, Department of Physics, Germany

4. LMU München, Department of Medical Physics, Germany

5. King Saud University, Department of Physics and Astronomy, Ryiadh, Saudi Arabia

6. Università degli Studi di Milano, Department of Physics, Italy

7. Delft University of Technology, Netherlands

Abstract

AbstractA Compton camera prototype for ion beam range monitoring via prompt (< 1 ns) gamma detection in hadron therapy is being developed and characterized at the Medical Physics Department of LMU Munich. The system consists of a large (50x50x30 mm3) monolithic LaBr3(Ce) scintillation crystal as absorber component to detect the multi-MeV Compton scattered photons, together with a stack of 6 double-sided silicon strip detectors (DSSSD) acting as scatterer component. Key ingredient of the γ-source reconstruction is the determination of the γ-ray interaction position in the scintillator, which is read out by a 256-fold segmented multi-anode photomultiplier tube (PMT). From simulations an angular resolution of about 1.5o for the photon source reconstruction can be expected for the energy range around 3 – 5 MeV, provided that a spatial resolution of 3 mm can be reached in the absorbing scintillator [1]. Therefore, particular effort was dedicated to characterize this latter property as a function of the γ-ray energy. Intense, tightly collimated 137Cs and 60Co photon sources were used for 2D irradiation scans (step size 0.5 mm) as prerequisite for studying the performance of the “k-Nearest-Neighbors” algorithm developed at TU Delft [2] (together with its variant ”Categorical Average Pattern”, CAP) and extending its applicability into the energy range beyond the original 511 keV. In this paper we present our most recent interaction position analysis in the absorbing scintillator, leading to a considerably improved value for the spatial resolution: systematic studies were performed as a function of the k-NN parameters and the PMT segmentation. A trend of improving spatial resolution with increasing photon energy was confirmed, resulting in the realization of the presently optimum spatial resolution of 2.9(1) mm @1.3 MeV, thus reaching the design specifications of the Compton camera absorber. The specification goal was reached also for a reduced PMT segmentation of 8x8 anode segments (each with 6x6 mm2 active area), thus allowing to reduce the complexity of the signal processing while preserving the performance.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering

Cited by 11 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3