Influence of pedicle screw thread width and recovery time after surgery on fixation strength

Author:

Makaram Harikrishna1,Swaminathan Ramakrishnan1

Affiliation:

1. Indian Institute of Technology Madras, Chennai , India

Abstract

Abstract Introduction: Pedicle screw fixation systems are widely used for treatment of various spinal pathologies, including spinal stenosis, scoliosis, spinal deformities and fractures. Stress shielding is considered to be a major factor contributing to insufficient fixation strength, leading to screw loosening. In this study, the influence of pedicle screw thread width on the displacement of pedicle screw and stress transfer is analyzed using 2-Dimensional axisymmetric finite element (FE) model. Methods: FE model consisting of cancellous and cortical bone, along with pedicle screw is developed for this study. The pedicle screw thread width is varied between 0.1 mm and 0.6 mm in steps of 0.1 mm, while the other geometric parameters, including screw half-angle, pitch, diameter, and length are kept constant. Three different contact conditions between screw and bone, such as frictionless, frictional, and bonded are considered to simulate hours, days, and months after surgery, respectively. The material properties and boundary conditions are applied based on previous studies. An axial force of 80 N is applied on the screw head to simulate axial pull-out test. Results: Similar patterns of stress distribution are observed for all screw models, with high stress concentration above the first thread. The highest displacement in screw is observed shortly after surgery, while the highest displacement in cancellous and cortical bone is observed few days and months after the surgery, respectively. The average von Mises stress in screw decreases with increase in thread width for all contact conditions. In few hours/days after the surgery, stress transfer parameter increases with increase in thread width, up to a thread width of 0.5 mm and then decreases. The changes in stress transfer parameter are negligible few months after the surgery. Conclusion: This study highlights the influence of thread width on displacement and stress transferred to the bone, at different durations after the surgery. It is observed that a thread width of 0.5 mm exhibits the highest stress transfer, leading to reduced stress shielding and improved bone remodeling. It appears that this study might aid in developing better pedicle screws for the treatment of various spinal pathologies.

Publisher

Walter de Gruyter GmbH

Subject

Biomedical Engineering

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Experimental Analysis of Stress Shielding Effects in Screw Spacers Placed in Porcine Spinal Tissue;Journal of Functional Biomaterials;2024-08-22

2. Biomechanical Study of Fused Lumbar Spine Comparing Ti and PEEK as Connecting Rod Considering Bone Degeneracy Using FEA;2022 29th National and 7th International Iranian Conference on Biomedical Engineering (ICBME);2022-12-21

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3