Author:
Pastor Ferenc T.,Dejmková Hana,Zima Jiří,Barek Jiří
Abstract
The possibility of determination of chloramphenicol by differential pulse voltammetry at four different carbon paste electrodes, in the full pH range (2–12) of Britton–Robinson (BR) buffer was investigated. Electrodes were prepared by mixing spectroscopic graphite powder or glassy carbon microbeads with mineral oil (Nujol) or tricresyl phosphate. Under optimal conditions (BR buffer pH 12, the electrode prepared from glassy carbon microbeads and tricresyl phosphate), linear calibration graph was obtained only in 10–5 M chloramphenicol concentration range. Determination of lower concentrations of chloramphenicol was complicated by irreproducible peak of oxygen from the carbon paste which overlapped with peak of chloramphenicol. Addition of sodium sulfite removed the oxygen peak without influence on the peak of chloramphenicol. Under optimal conditions (electrode paste made from glassy carbon microbeads, BR buffer pH 10 and 0.5 M sodium sulfite), straight calibration line was obtained in the 10–6 and 10–5 M chloramphenicol concentration range. Limit of determination was 5 × 10–7 mol/l.
Publisher
Institute of Organic Chemistry & Biochemistry
Cited by
5 articles.
订阅此论文施引文献