Abstract
Background/Aims: Endoscopic ultrasound (EUS)-guided hepaticogastrostomy (EUS-HGS) performed at the intrahepatic bile duct segment 3 (B3) is widely used for biliary drainage. Although performing post-puncture procedures is easier in the intrahepatic bile duct segment 2 (B2) when using a conventional oblique-viewing (OV) EUS scope, this method may cause transesophageal puncture and severe adverse events. We evaluated the safety and efficacy of B2 puncture using a novel OV-EUS scope.Methods: In this single-center retrospective study, we prospectively enrolled and collected data from 45 patients who consecutively underwent EUS-HGS procedures with a novel OV-EUS scope between September 2021 and December 2022 at our cancer center.Results: The technical success rates of B2-EUS-HGS and EUS-HGS were 93.3% (42/45) and 97.8% (44/45), respectively. The early adverse event rate was 8.9% (4/45) with no cases of scope changes or transesophageal punctures. The median procedure time was 13 minutes (range, 5–30).Conclusions: B2-EUS-HGS can be performed safely with the novel EG-740UT (Fujifilm) OV-scope without transesophageal puncture and with a high success rate. B2-EUS-HGS using this novel OV scope may be the preferred strategy for EUS-HGS.
Publisher
The Korean Society of Gastrointestinal Endoscopy
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献