Is a flared stent effective for decreasing stent migration in canine prostatic urethra?

Author:

Woo Chul-Woong1,Song Ho-Young1,Yoon Chang Jin1,Kim Tae-Hyung1,Kim Eun-Young1,Kim Jin Hyoung1,Shin Ji Hoon1

Affiliation:

1. Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, Republic of Korea; Department of Radiology, Seoul National University Bundang Hospital, Seongnam, Republic of Korea

Abstract

Background: Prostatic urethral stents are effective in relieving obstructions caused by benign prostatic hyperplasia (BPH). However, migration of these stents occurs frequently. Purpose: To evaluate the efficacy of a flared stent for decreasing the migration rate in comparison with a straight stent in a canine prostatic urethral model. Material and Methods: The flared stent (15 mm in diameter and 20 mm in length) was flared up to 19 mm at both ends to prevent migration. A straight stent with the same size was straight without flaring. Both stents were made of a nitinol wire filament and covered with an expanded polytetrafluoroethylene (ePTFE) membrane. The flared stent was inserted in the prostatic urethra of 10 dogs (group 1) and the straight stent in the prostatic urethra of 12 dogs (group 2). Follow-up retrograde urethrography (RUG) was performed 1, 4, and 8 weeks after stent placement. Fisher's exact test was used to compare the migration rate between the two groups. When stent migration occurred during the follow-up period, the same type of stent was inserted again. Results: Three of 10 (30.0%) flared stents migrated into the urinary bladder between 1 week and 4 weeks after placement. Seven of 12 (58.3%) straight stents migrated into the urinary bladder 1 ( n = 3), 4 ( n = 1), and 8 ( n = 3) weeks following placement. The flared stent group showed lower migration rate than the straight stent group, but the migration rate did not reach statistical significance ( P = 0.231). After the second stent placement, one of three (33.3%) flared stents migrated again after 4 weeks and two of seven straight stents (28.6%) after 1 week. Conclusion: Although the migration rate was not statistically significant, the flaring of the stent seemed to reduce the migration rate in comparison with straight stents in a canine prostatic urethral model. However, the migration rate of the flared stent was still high, and further developments are required to decrease migration rate.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

Cited by 11 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司