Soft-copy reading in digital mammography of mass: diagnostic performance of a 5-megapixel cathode ray tube monitor versus a 3-megapixel liquid crystal display monitor in a diagnostic setting

Author:

Uematsu T.1,Kasami M.1

Affiliation:

1. Breast Imaging Section and Department of Pathology, Shizuoka Cancer Center Hospital, Shizuoka, Japan

Abstract

Background: Liquid crystal display (LCD) monitors and cathode ray tube (CRT) monitors are currently the two most common types used in digital mammography systems. The appropriate selection of a monitor is very important and requires balancing the monitor's performance and its cost. A previous study of soft-copy reading in digital mammography of microcalcifications showed that 3-megapixel (M) LCD monitors were similar in diagnostic performance to 5M CRT monitors in a diagnostic setting. Purpose: To compare 5M CRT monitors with 3M LCD monitors for soft-copy reading of digital mammography of a mass in a diagnostic setting. Material and Methods: Seventy mass lesions having undergone either breast biopsies or definitive surgery (46 malignant and 24 benign) and 30 normal cases were recruited into the study. The median size of the lesions was 16 mm (range 7–20 mm). The digital mammograms in the 100-case set were assigned to two blocks, block A (50 cases) and block B (50 cases), for 5M CRT and 3M LCD monitors. A single radiologist read all 100 cases with both types of monitors, starting with the images in block A on the 5M CRT monitors and then the images in block B on the 3M LCD monitors. The radiologist analyzed the soft-copy images on 5M CRT and 3M LCD monitors with 5 months between the interpretations to reduce the effects of learning and memory. Again, the reader started with the images in block A on the 3M LCD monitors and then read the images in block B on the 5M CRT monitors. A five-point rating scale for the probability of malignancy was used for interpreting the soft-copy mammograms. The mass descriptor was scored on a six-point scale. Breast density was scored on a four-point scale. The positive predictive value (PPV) and negative predictive value (NPV) were calculated based on the criteria of the Breast Imaging Reporting and Data System. The interpretation time was also measured. Results: No significant difference was observed in the probability of malignancy ( P=1), mass descriptor ( P=0.317), and breast density ( P=0.739). The PPV and NPV of soft-copy reading on the 5M CRT monitors were 91% (42/46) and 94% (51/54), respectively, identical to the results using 3M LCD monitors. The total interpretation time averaged 62 s for the 5M CRT monitors and 60 s for the 3M LCD monitors ( P<0.0001). Conclusion: Soft-copy reading of a digital mammography of mass with 3M LCD monitors was similar in diagnostic performance to 5M CRT monitors in this study. On the basis of the results of this and a previous study, 3M LCD monitors can replace 5M CRT monitors without any loss in the ability to diagnose digital mammograms.

Publisher

SAGE Publications

Subject

Radiology Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3