Magnetic resonance imaging of hyaline cartilage defects at 1.5T and 3.0T: comparison of medium T2‐weighted fast spin echo, T1‐weighted two‐dimensional and three‐dimensional gradient echo pulse sequences
-
Published:2005-02
Issue:1
Volume:46
Page:67-73
-
ISSN:0284-1851
-
Container-title:Acta Radiologica
-
language:en
-
Short-container-title:Acta Radiol
Author:
Fischbach F.1, Bruhn H.1, Unterhauser F.1, Ricke J.1, Wieners G.1, Felix R.1, Weiler A.1, Schröder R. J.1
Affiliation:
1. Department of Radiology and Department of Traumatic and Reconstructive Surgery, Charité, Campus Virchow-Klinikum, Humboldt-University Medical School, Berlin, Germany
Abstract
Purpose: To evaluate and compare the diagnostic accuracy of appropriate magnetic resonance (MR) sequences in the detection of cartilage lesions at 1.5T and 3.0T. Material and Methods: Twelve chondral defects of varying depths, widths, and locations were created in the retropatellar hyaline cartilage in six sheep cadaver limbs. Axial images employing three fat‐suppressed imaging sequences – (1) a T2‐weighted fast spin‐echo (FSE) sequence, (2) a two‐dimensional (2D) and (3) three‐dimensional (3D) gradient‐echo (GE) sequence at 1.5T and 3.0T using an extremity quadrature coil – were evaluated by three experienced radiologists. Statistical analysis of the results consisted of receiver operating characteristics (ROC) and significant testing using the bivariate chi‐square test. In addition, signal‐to‐noise ratios (SNR) and contrast‐to‐noise ratios (CNR) were evaluated with significance testing using the Wilcoxon test. Results: The 3D GE sequence compared favorably with other sequences at 3.0T and 1.5T (Az = 0.88 at 3.0T and Az = 0.85 at 1.5T) missing only one small grade 2 lesion. 2D GE imaging was inferior to 3D imaging at both field strengths ( P<0.05) in general. However, compared to 1.5T, lesion detectability was improved at the higher magnetic field of 3.0T (Az = 0.81 and 0.73 at 3.0T and 1.5T, respectively). FSE images showed significantly inferior sensitivity and less anatomical detail compared to the GE sequences at both field strengths (Az = 0.64 and 0.72 at 3.0T and 1.5T, respectively; P<0.05). However, compared to 1.5T, lesion detectability SNR and CNR values were superior in all sequences tested at 3.0T. Conclusion: MRI at 3.0T improves SNR and CNR significantly in the most common sequences for cartilage MRI, resulting in an improvement in chondral lesion detection. GE imaging therefore allows resolution to be increased in an acceptable time manner for patient comfort, and the 3D GE fat‐suppressed sequence at 3.0T appears to be best suited for cartilage imaging in a clinical setting.
Publisher
SAGE Publications
Subject
Radiology, Nuclear Medicine and imaging,General Medicine,Radiological and Ultrasound Technology
Reference17 articles.
1. Chandnani, V. P, Ho, C, Chu, P, Trudell, D, Resnick, D. Knee hyaline cartilage evaluated with MR imaging: a cadaveric study involving multiple imaging sequences and intraarticular injection of gadolinium and saline solution. Radiology, 1991; 178: 557–61 2. Disler, D. G, McCauley, T. R, Kelman, C. G, et al, . Fat‐suppressed three‐dimensional spoiled gradient‐echo MR imaging of hyaline cartilage defects in the knee: comparison with standard MR imaging and arthroscopy. Am J Roentgenol, 1996; 167: 127–32 3. Disler, D. G, Recht, M. P, McCauley, T. R. MR imaging of articular cartilage. Skeletal Radiol, 2000; 29: 367–77 4. Mlynarik, V, Degrassi, A, Toffanin, R, et al, . Investigation of laminar appearance of articular appearance by means of magnetic resonance microscopy. Magn Reson Imaging, 1996; 14: 435–42 5. Mosher, T, Smith, H, Dardzinski, B. J, et al, . MR imaging and T2 mapping of femoral cartilage: in vivo determination of the magic angle effect. Am J Roentgenol, 2001; 177: 665–9
Cited by
52 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|