Abstract
Abstract
Purpose
The aim of this study was to compare the enucleation performances of four different types of laser devices in an ex-vivo experiment: a novel, pulsed Tm:YAG solid-state laser evaluation model (p-Tm:YAG), chopped thulium fibre laser (TFL), low-power Ho:YAG laser (LP-Ho:YAG), and a high-power Ho:YAG laser (HP-Ho:YAG).
Methods
Our primary aim was to endoscopically separate the fascial layers of a porcine belly using laser fibres within a time period of 60 s. The size of a “tissue pocket” was assessed numerically. The enucleation characteristics reflecting the surgeon’s experience were evaluated via the NASA Task Load Index (TLX) questionnaire and a questionnaire based on Likert scale.
Results
HP-Ho:YAG achieved with the available laser settings the largest overall “tissue pocket” (31.5 cm2) followed by p-Tm:YAG (15 cm2), TFL (12 cm2), and LP-Ho:YAG (6 cm2). The coagulation performances of p-Tm:YAG and TFL were rated the best. In the performance evaluation by the Likert questionnaire, HP-Ho:YAG (average score of 4.06) was rated highest, followed by p-Tm:YAG (3.94), TFL (3.38), and LP-Ho:YAG (3.25). The evaluation of the NASA-TLX performance questionnaire revealed average scores for HP-Ho:YAG, LP-Ho:YAG, TFL and p-Tm:YAG of 4.38, 4.09, 3.92 and 3.90, respectively.
Conclusion
We are the first to compare different laser devices and settings in an ex-vivo study. We found that the surgeons were most satisfied with the HP-Ho:YAG laser device, followed by the p-Tm:YAG. These findings could be highly relevant for future research and for the practical utilisation of laser systems in endourology.
Funder
Dornier MedTech Laser GmbH
Albert-Ludwigs-Universität Freiburg im Breisgau
Publisher
Springer Science and Business Media LLC
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献